Skip to main content
. 2018 Jun 28;13(1):55–59. doi: 10.1177/1932296818783783

Table 2.

Precision Data for Actrapid and Insulatard.

Result of analysis (units/mL), Assay
Actrapid 100 Units/mL Mean 99.4632
% RSD (%) Repeatability 0.1361
Intermediate precision 0.2529
95% confidence interval % RSD Repeatability 0.0877-0.2997
Intermediate precision 0.1659-0.5267
Insulatard 100 units/mL Mean 99.7729
% RSD (%) Repeatability 0.2332
Intermediate precision 0.3762
95% confidence interval % RSD Repeatability 0.1502-0.5134
Intermediate precision 0.2505-0.7499