Table 2.

Key Performance Measures Tested in Human Factors Validation Study.

Key performance measures N = 30
Set up tasks
Able to setup PAQ 30/30 (100%)
Able to fill PAQ with insulin 30/30 (100%)
Able to prime (remove air) from fluid path 30/30 (100%)
Able to apply and adhere device to skin 30/30 (100%)
Able to deploy and insert cannula 30/30 (100%)
Bolus tasks
Able to deliver 8 or 16 U bolus dose (no distraction) 30/30 (100%)
Able to deliver 8 or 16 U bolus dose (with distraction) 30/30 (100%)
Messenger button
Able to identify and press Messenger button 30/30 (100%)
Able to identify the number of vibrations emitted 30/30 (100%)
Able to interpret the meaning of the vibrations, and take appropriate action 30/30 (100%)
Replace the reservoir after 3 days
Able to remove the PAQ, disengage the Messenger component, and replace the reservoir 30/30 (100%)