Table 2:
Step | Problem | Possible reason | Solution |
---|---|---|---|
7 | Wells are not properly positioned and walls are not perpendicular | Biopsy punches were not properly used | Print a schematic of the design and use it as a guide. Pay attention to applying pressure perpendicularly to the surface |
8 | The 1.5 mm diameter connecting channel is difficult to punch | There is not sufficient structural integrity to pierce the PDMS | Place the cored 8-mm plug back inside the well to absorb the pressure of the 1.5-mm punch |
28 | Germaria are bunched up with egg chambers in the muscle sheath | The muscle sheath was intact. It should be broken/ ripped near the stage 7 egg chamber. | During dissection pinch off stage 7/8 and older egg chambers from the ovariole with forceps. |
41 | Images become darker over time in transmitted light channel Imaging software cannot handle the large dataset being acquired |
An air bubble formed just below the chamber lid Images are being saved to memory |
Make two 1-mm holes in the PDMS lid above the imaging well Use settings in imaging software to autosave to hard drive instead |