Published in final edited form as: Nat Protoc. 2018 Nov;13(11):2601–2614. doi: 10.1038/s41596-018-0054-1

Table 2:

Troubleshooting Table

Step Problem Possible reason Solution
7 Wells are not properly positioned and walls are not perpendicular Biopsy punches were not properly used Print a schematic of the design and use it as a guide. Pay attention to applying pressure perpendicularly to the surface
8 The 1.5 mm diameter connecting channel is difficult to punch There is not sufficient structural integrity to pierce the PDMS Place the cored 8-mm plug back inside the well to absorb the pressure of the 1.5-mm punch
28 Germaria are bunched up with egg chambers in the muscle sheath The muscle sheath was intact. It should be broken/ ripped near the stage 7 egg chamber. During dissection pinch off stage 7/8 and older egg chambers from the ovariole with forceps.
41 Images become darker over time in transmitted light channel
Imaging software cannot handle the large dataset being acquired
An air bubble formed just below the chamber lid
Images are being saved to memory
Make two 1-mm holes in the PDMS lid above the imaging well
Use settings in imaging software to autosave to hard drive instead