Table 3. Euro-CMR Scores for Image Acquired by Segmented Cine and Those by Real-Time Cine.

Segmented Cine (Group 1, n = 33/Group 2, n = 15) Real-Time Cine (Group 1, n = 33/Group 2, n = 15)
0 1 2 3 0 1 2 3
Wrap-around 33/15 -/- -/- -/- 33/15 -/- -/- -/-
Respiratory ghost 33/10 -/4 -/- -/1 33/15 -/- -/- -/-
Cardiac ghost 33/15 -/- -/- -/- 24/11 8/4 1/- -/-
Blurring/mis-triggering 33/3 -/3 -/5 -/4 33/14 -/1 -/- -/-
Metallic artifact 33/15 -/- -/- -/- 33/15 -/- -/- -/-
Shimming artifact 28/14 3/- 1/1 1/- 33/15 -/- -/- -/-
Total score, median (IQR) 0.3 ± 0.7; 0 (0, 0)/2.3 ± 1.2; 2 (1, 3) 0.3 ± 0.5; 0 (0, 1)/0.4 ± 0.5; 0 (0, 0)

Group 1, patients without arrhythmia or difficulty in breath-holding; Group 2, patients with arrhythmia and/or difficulty in breath-holding. Euro-CMR = European CMR Registry, IQR = interquartile range