Skip to main content
. 2018 Oct 17;8(4):93. doi: 10.3390/bios8040093

Table 2.

Influence on overall MARD of different signal processing of day 1 data to correct for bias and lag.

Signal Processing Symbol MARD
(%)
MAD
(mmol/L)
PEG zone A/A+B
(%)
None (raw FGM scans vs SMBG) yFGM 12.3 1.0 81.7/100
Bias corrected, multipoint yFGM,Bcorr 9.2 0.8 91.2/100
Bias corrected, 1-point yFGM,Bcorr1p 11.4 0.9 83.0/100
Bias corrected, 2-point yFGM,Bcorr2p 9.7 0.7 87.7/100
Bias and lag corrected, multipoint yFGM,BLcorr 6.6 0.5 97/100
Only lag corrected, multipoint yFGM,Lcorr 11.7 0.9 81.5/100
Only smoothed, multipoint yFGM,smoothed 11.9 1.0 82.2/100