HIV seroconversion panel |
HIV-1/2 antibodies were detected 0.125 specimens earlier compared to the reference assay |
HIV-1/2 antibodies were detected 0.25 specimens later compared to the reference assay |
HIV-1/2 antibodies were detected 0.125 specimens later compared to the reference assay |
HIV-1/2 antibodies were detected at the same time as the reference assay |
HIV mixed titre panels |
All were correctly detected |
88% were correctly detected (22/25; one false reactive, one false non-reactive and one indeterminate) |
96% were correctly detected (24/25; one false non-reactive) |
All were correctly detected |
WHO HIV Reference preparations |
All were correctly detected, with the exception of the Group O specimen (indeterminate) |
All were correctly detected |
All were correctly detected |
All were correctly detected |
TP seroconversion panels |
TP antibodies were detected 2 specimens earlier compared to the reference assay |
TP antibodies were detected at the same time compared to the reference assay |
TP antibodies were detected 2 specimens later compared to the reference assay |
TP antibodies were detected 1 specimen earlier compared to the reference assay |
TP mixed titre panels |
All were correctly detected |
All were correctly detected |
All were correctly detected |
All were correctly detected |
WHO TP Reference preparations |
All were correctly detected |
All were correctly detected |
All were correctly detected |
All were correctly detected |