Table 2.
Task | Mean blink rate for each task among ventilators | p | Post hoc tests with Bonferroni correction | p | ||||
---|---|---|---|---|---|---|---|---|
| ||||||||
Evital 4 | Servo I | Boaray 5000D | ||||||
|
|
|||||||
Mean±SD | Mean±SD | Mean±SD | Mr1 | Mr2 | MD (Mr1–Mr2) | |||
Task1: start the ventilator (second) | 0.117±0.170 | 0.101±0.019 | 0.101±0.009 | p<.01 | Servo I | Evital 4 | −0.016 | 0.01 |
| ||||||||
Boaray 5000D | Evital 4 | −0.016 | 0.02 | |||||
| ||||||||
Servo I | Boaray 5000D | 0.000 | 1.00 | |||||
| ||||||||
Task2: mode and setting parameters recognition (second) | 0.097±0.014 | 0.123±0.026 | 0.111±0.026 | 0.02 | Servo I | Evital 4 | 0.026 | 0.02 |
| ||||||||
Boaray 5000D | Evital 4 | 0.014 | 0.40 | |||||
| ||||||||
Servo I | Boaray 5000D | 0.012 | 0.65 | |||||
| ||||||||
Task3: ventilator monitoring values recognition (second) | 0.091±0.010 | 0.123±0.010 | 0.111±0.020 | p<.01 | Servo I | Evital 4 | 0.032 | p<.01 |
| ||||||||
Boaray 5000D | Evital 4 | 0.020 | 0.01 | |||||
| ||||||||
Servo I | Boaray 5000D | 0.012 | 1.00 | |||||
| ||||||||
Task4: ventilator setting parameters modification (second) | 0.104±0.018 | 0.123±0.034 | 0.114±0.027 | 0.45 | ||||
| ||||||||
Task5: ventilator mode modification (second) | 0.093±0.014 | 0.118±0.021 | 0.112±0.018 | 0.01 | Servo I | Evital 4 | 0.025 | 0.01 |
| ||||||||
Boaray 5000D | Evital 4 | 0.019 | 0.28 | |||||
| ||||||||
Servo I | Boaray 5000D | 0.006 | 0.75 | |||||
| ||||||||
Task6: alarm parameter recognition and resetting (second) | 0.094±0.013 | 0.109±0.019 | 0.118±0.022 | 0.03 | Servo I | Evital 4 | 0.015 | 0.23 |
| ||||||||
Boaray 5000D | Evital 4 | 0.024 | 0.04 | |||||
| ||||||||
Servo I | Boaray 5000D | −0.009 | 1.00 | |||||
| ||||||||
Task7: respond to alarm (second) | 0.097±0.023 | 0.112±0.025 | 0.104±0.023 | 0.29 | ||||
| ||||||||
Average blink duration of all tasks (second) | 0.099±0.009 | 0.116±0.013 | 0.111±0.009 | p<.01 | Servo I | Evital 4 | 0.017 | 0.01 |
| ||||||||
Boaray 5000D | Evital 4 | 0.005 | 0.01 | |||||
| ||||||||
Servo I | Boaray 5000D | 0.012 | 1.00 |
Positive MD values representing Mr1 lower workload than Mr2.