Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2018 May 16;1090:22–35. doi: 10.1016/j.jchromb.2018.05.008

Table 3.

Intra- and inter-assay accuracy and precision of DHT, α-diol, and β-diol in pooled mouse sera (n = 5).

Intra-Assaya
DHT α-diol β-diol
[XN] (ng/ml) [X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
[X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
[X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
0.150 0.14 ± 0.01 7 −7 0.15 ± 0.01 7 0 0.15 ± 0.01 7 0
2.50 2.30 ± 0.02 1 −8 2.43 ± 0.03 1 −3 2.51 ± 0.07 3 0
40.0 37 ± 1 3 −8 38 ± 2 5 −5 39 ± 1 3 −3
Inter-Assayb
DHT α-diol β-diol
[XN] (ng/ml) [X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
[X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
[X¯] ± SD
(ng/ml)
CV
(%)
RE
(%)
0.150 0.14 ± 0.01 7 −7 0.15 ± 0.01 7 0 0.16 ± 0.01 6 7
2.50 2.4 ± 0.1 4 −4 2.5 ± 0.1 4 0 2.47 ± 0.05 2 −1
40.0 36.7 ± 0.4 1 −8 38.6 ± 0.5 1 −4 39.8 ± 0.7 2 −1
a

Measured by five replicate measurements of each QC sample within a validation batch.

b

Measured by five parallel measurements of five identical QC samples at each concentration over five validation batches.