Solid state form |
X-ray powder diffraction (XRPD) |
Diffraction of x-rays from lattice planes |
Polymorphic form (unique diffraction peaks), amorphous form (no peaks) |
Diffractogram, qualitative and quantitative (degree of crystallinity) |
Hot stage XRPD to analyse solid state form as a function of temperature |
Powder, paste or slurry form, several sample presentation setups possible, amount required depends on setup |
Anisotropic particle shape leads to preferred orientation effects (change in relative intensities of diffraction peaks) |
[17] |
Peak broadening can occur as crystal lattice size decreases within nanoscale range |
Differential scanning calorimetry (DSC) |
Change in heat flow due to sample changes during heat/cooling |
Polymorphic form (melting temperature, crystallisation temperature) amorphous form (glass transition temperature), crystallinity (enthalpy of fusion, enthalpy of crystallisation, heat capacity change at glass transition temperature) |
Thermogram, qualitative and quantitative |
Modulated temperature DSC to separate overlapping irreversible and reversible thermal events, ultrafast heating |
Powder form, few milligrams |
Destructive. Results will be different with open or closed (hermetically sealed) pans |
[18] |
Infrared (IR) spectroscopy (mid-IR spectroscopy) |
Change in dipole moment during molecular vibrations |
Polymorphic form (peak shifts and relative intensities), crystallinity (broadening of bands, peak shifts and relative intensities) |
Spectrum, qualitative and quantitative, suitable for multivariate analysis |
Diffuse reflectance IR (DRIFTS), attenuated total reflection (ATR), microscope |
Powder or tablet form, depends on sampling setup, few milligrams. Wet samples usually problematic. |
Sample preparation/measurement can involve pressure which can induce solid state transformations |
[19,20] |
Raman spectroscopy |
Change in polarisability during molecular vibrations |
Polymorphic form, crystallinity |
Spectrum, qualitative and quantitative, suitable for multivariate analysis |
Various sample holders (within spectrometer, sampling probes, microscope) |
Powder or suspension, few milligrams (usually). Fluorescent samples are problematic. |
Sample heating can be problematic. Samples can be in aqueous medium. |
[20,21,22] |
Size and morphology |
Dynamic light scattering (photon correlation spectroscopy) |
Fluctuation of Rayleigh scattering of light associated with Brownian motion of nanoparticles |
Particle size, particle size distribution |
Particle size distribution (number based mean particle (hydrodynamic) size (Z-average), polydispersity index), quantitative |
|
Suspension with suitable concentration |
Suitable only for particles in nanometre size range |
[23] |
Viscosity of suspension and temperature affect results |
Scanning electron microscopy (SEM) |
Backscattering of electrons |
Topographical information about particles |
Scanning electron micrograph, particle morphology, size |
Elemental analysis |
Dry sample mounted on stage condition setup (vacuum), microgram requirement |
Sample preparation destructive |
[15] |
Transmission electron microscopy |
Transmission of electrons |
Density information |
Transmission electron micrograph, morphology of cross sections, stabilizer- nanocrystal interaction |
|
Embedded cross section preparation, microgram requirement |
Sample preparation destructive |
[24] |
Surface properties |
Zeta-potential |
Dynamic electrophoretic mobility under electric field |
Surface charge (zeta potential) |
Zeta potential, quantitative |
|
Suspension with suitable concentration |
|
[25] |
Surface plasmon resonance (SPR) |
Changes in refractive index in the vicinity of a planar sensor surface |
Surface adsorption |
Spectrum, interaction between stabiliser drug crystals, qualitative and quantitative |
|
Substrate on planar surface sensor required (not direct measurement of nanocrystals) |
Careful sample preparation required |
[26] |
Drug delivery |
Dissolution testing |
Dissolved drug analysed over time, usually using UV spectroscopy or HPLC |
Dissolution profile |
Solution concentration vs time |
Paddle, flow through cell (with/without membrane insert), pharmacopeial/non pharmacopeial |
|
Separating nanocrystals from dissolution medium can be problematic |
[14] |
Fluorescence microscopy |
Fluorescence by endogenous or added fluorophores |
Localization of nanocrystals in relation to cells and tissues |
Fluorescence (and nanocrystal) image |
One or two photon (two photon fluorescence offers inherent confocality, sub-micron spatial resolution, deeper penetration in tissues) fluorescence |
Non-fluorescent nanocrystals require fluorphore to physically entrapped into nanocrystals |
Entrapment and leakage of fluorophore can be difficult or problematic |
[17] |
Non-linear Raman microscopy |
Change in polarisability during molecular vibrations. |
Label free localisation of particles |
Intensity of CARS shift (narrow band) or spectrum, (multiplex or broad band). Most commonly qualitative. 2D or 3D images. |
Can be dry or aqueous suspension, in cell cultures or tissue samples |
Coloured and two-photon fluorescent samples can interfere with signal. Can be coupled with other nonlinear phenomena such as second harmonic generation or two photon electronic fluorescence |
Label free. Optimal lateral spatial resolution approximately 300–400 nm. |
[21,27] |