Skip to main content
. 2018 Dec 7;20(1):184–193. doi: 10.1002/acm2.12507

Table 1.

Photon open‐field point‐dose verification

Model Energy SSD (cm) Field size (cm²) Depth (cm) % (Meas‐Calc)/average
MC 6 MV 90 10 × 10 20.0 0.2
100 40 × 5 1.5 −0.7
6 FFF 90 2 × 2 5.0 0.6
100 40 × 40 10.0 0.0
10 MV 90 30 × 30 15.0 −0.5
100 5 × 20 2.2 −1.6
10 FFF 90 30 × 30 2.4 0.1
100 5 × 5 5.0 −0.7
18 MV 90 10 × 10 20.0 0.4
100 5 × 40 3.0 −2.1
CC 6 MV 90 3 × 3 5.0 0.2
100 20 × 20 1.5 −1.4
10 MV 90 30 × 30 10.0 −0.9
100 2 × 2 2.2 −2.5
18 MV 90 5 × 5 10.0 0.7
100 40 × 40 3.0 −0.4