Published in final edited form as: ASAIO J. 2019 Jul;65(5):465–472. doi: 10.1097/MAT.0000000000000835

Table 1.

Measured aortic blood pressure (BP; units of mmHg and represented as systolic/diastolic) and heart rate (HR; units of beats/minute). All numbers are average values determined by selecting a range of cardiac cycles that correspond to the physiologic interventions (i.e., baseline, increased preload, and increased afterload) and pumping mode (i.e., continuous and pulsed).

Baseline Increased Preload Increased Afterload

Continuous Pulsed Continuous Pulsed Continuous Pulsed

BP HR BP HR BP HR BP HR BP HR BP HR

Centrifugal 1 63/38 136 72–35 136 67/37 145 77/34 146 90/44 153 96/40 154
2 69/35 96 69/27 98 68/25 86 72/19 77 84/41 140 87/34 142

Axial 1 59/43 89 50/40 89 64/38 138 64/38 138 73/46 149 71/45 149
2 76/42 132 76/41 132 76/28 115 75/28 115 86/44 149 85/43 148