Table 3.

Optimal SediVue (SW1.0.1.3) thresholds for the detection of formed elements in urine based on ROC analysis with corresponding sensitivities and specificities, as compared to manual microscopy

Optimal threshold Sensitivity (95% CI) Specificity (95% CI)
RBC 6.5/hpf 88.3% (82.5%‐92.7%) 92.8% (89.6%‐95.2%)
WBC 6.5/hpf 83.3% (75.7%‐89.4%) 92.6% (89.6%‐94.9%)
sqEPI 0.2/hpf 75.0% (53.3%‐90.2%) 90.3% (87.4%‐92.8%)
nsEPI 0.9/hpf 75.0% (61.6%‐85.6%) 85.9% (82.4%‐88.9%)
STR 1.0/hpf 90.6% (84.1%‐95.0%) 84.4% (80.5%‐87.8%)
CaOx Di 0.3/hpf 84.6% (71.9%‐93.1%) 97.9% (96.2%‐99.0%)

Abbreviations: CaOx Di, calcium oxalate dihydrate crystals; CI, confidence intervals; hpf, high power field; nsEPI, non‐squamous epithelial cells; RBC, red blood cells; sqEPI, squamous epithelial cells; STR, struvite crystals; WBC, white blood cells.