Skip to main content
. 2019 Jan 8;20(1):201. doi: 10.3390/ijms20010201

Table 1.

Peak capacity (nc), selectivity (α), retention factor (k), and resolution (Rs), calculated for the EVOO hydrophilic phenolic extract, for the peak pairs HT–T (hidroxytyrosol–tyrosol) and OAIa–OAIb (oleuropein aglycone isomers “a” and “b”), for the three studied columns (Kinetex biphenyl, LiChrospher C18, and Spherisorb C18).

Column nc α k Rs
HT–T OAIa–OAIb HT HT–T OAa–OAb
Kinetex biphenyl 277 1.43 1.045 1.46 5.60 2.68
LiChrospher C18 192 1.38 1.024 1.81 5.39 1.12
Spherisorb C18 179 1.37 1.0 1.94 3.91 0