Published in final edited form as: J Ultrasound Med. 2018 Jul 19;38(2):307–319. doi: 10.1002/jum.14690

Table 1.

Ultrasound System Settings for Imaging Near and Far Zone Vessels in a Grayscale Small Parts Phantom.

System Identification
Code
Transducer
&
Transmission
Frequency
(MHz)
Dynamic Range
(dB) or
Compression
Most Linear
Grayscale
Map(*)
MI
(Near &
Far)
Mean Optimum
Gain Value per
Vessel
Near (Range), Far
(Range)
A L14-5/38 145 6 <0.80 66.8(58–77), 71.2(58–84)
Ultrasonix SonicTOUCH (BK Ultrasound, Peabody, MA, USA) H 10.0 <0.75
B 9L4 90 A 0.90 0.0(0-0), 0.0(0-0)
Acuson S2000 (Siemens Medical Solutions, Malvern, PA, USA) H 9.0 0.80
C L10-5 70 L 0.50 8.8(8–10), 15.2(12–18)
Siemens CV 70 (Siemens Medical Solutions, Malvern, PA, USA) 10.0 0.50
D 8L5 100 4 0.80 8.0(4–14), 8.6(5–13)
Acuson Sequoia (Siemens Medical Solutions, Malvern, PA, USA) 8.0 0.59
E L9-3 70 5 0.70 47.6(45–50), 48.6(46–50)
Philips iE33 (Philips Healthcare, Andover, MA, USA) Gen 0.90
F 9L4 90 A 1.3 0.0(0-0), 0.0(0-0)
Acuson S2000 (Siemens Medical Solutions, Malvern, PA, USA) H 8.0 1.1
G 9L 102 C 1.2 3.2(2–4), 3.2(2–4)
GE VIVID E9 (GE Healthcare, Waukesha, WI, USA) 4.0/8.0 1.2

MI = mechanical index

H = harmonic

*

System G clinical carotid grayscale map.