Table 2.

Respiratory measurements according to different techniques—retrospective phase

Number of patients with reference values
Manual-RR n-(%) 199/201 (99.0)
Ventilator-RR n-(%) 128/128 (100)
PPG-RR n-(%) 193/201 (96.0)
Measurements (c/min)
Reference RR Estimated RR-PPG R ICC p
22 ± 4 22 ± 5 0.955 0.931 < 0.0001
Reference RR under MV Estimated RR-PPG under MV
21.9 ± 5.4 21.8 ± 5.6 0.952 0.793 < 0.0001
 With VT > median  With VT > median
 20.1 ± 3.9  20.0 ± 4.1 0.947 0.969 < 0.0001
 With VT ≤ median  With VT ≤ median
 22.6 ± 6.1  22.5 ± 6.2 0.974 0.886 < 0.0001
Reference RR under SB Estimated RR-PPG under SB
23.2 ± 5.8 22.9 ± 5.9 0.940 0.963 < 0.0001
Reference RR without AF Estimated RR-PPG without AF
22.1 ± 5.6 22.0 ± 5.8 0.954 0.973 < 0.0001
Reference RR with AF Estimated RR-PPG with AF
20.6 ± 3.7 20.5 ± 4.0 0.916 0.689 < 0.0001

Values are provided as n-(%) or mean ± SD (b/mn), unless specified otherwise. R, Pearson’s correlation coefficient; ICC, intraclass correlation coefficient for average measurements, was assessed to evaluate the reliability of measurements; a p value < 0.01 was considered statistically significant for a bilateral test

Reference RR values are provided as the chronometric measurement measured over 30 s under spontaneous ventilation or as the ventilatory rate for patients under ventilatory assistance

RR, respiratory rate; PPG, plethysmographic measurement; Manual-RR, chronometric evaluation of the RR, over a 30-s period; Ventilator-RR, RR value provided by the ventilator in case of mechanical ventilation; PPG-RR, estimated respiratory rate using the plethysmographic curve analysis; MV, mechanical ventilation; SN, spontaneous breathing; AF, atrial fibrillation; VT, tidal volume (median tidal volume = 475 mL)