Table 2. RMSE Values of Subtracted CT Volumes in Test 1.
Conversion Pairs (Input, Target Image) | Original (HU) | Converted (HU) | Differences (%) |
---|---|---|---|
From smooth to sharp kernel | |||
B10f, B30f | 16.31 ± 2.04 | 6.00 ± 0.91 | −63.2 ± 2.9 |
B10f, B50f | 58.53 ± 11.02 | 33.08 ± 8.71 | −44.1 ± 5.2 |
B10f, B70f | 119.48 ± 28.10 | 84.18 ± 25.54 | −30.5 ± 5.4 |
B30f, B50f | 44.57 ± 9.64 | 18.29 ± 4.34 | −59.1 ± 1.8 |
B30f, B70f | 107.47 ± 27.29 | 52.95 ± 15.35 | −51.1 ± 2.0 |
B50f, B70f | 66.03 ± 18.32 | 22.82 ± 6.63 | −65.5 ± 1.9 |
From sharp to smooth kernel | |||
B30f, B10f | 16.31 ± 2.04 | 4.05 ± 0.36 | −75.0 ± 1.7 |
B50f, B10f | 58.53 ± 11.02 | 12.91 ± 2.65 | −77.9 ± 1.8 |
B50f, B30f | 44.57 ± 9.64 | 10.72 ± 2.37 | −75.8 ± 2.1 |
B70f, B10f | 119.48 ± 28.10 | 24.89 ± 6.55 | −79.1 ± 2.4 |
B70f, B30f | 107.47 ± 27.29 | 22.14 ± 5.98 | −79.2 ± 2.6 |
B70f, B50f | 66.03 ± 18.32 | 16.95 ± 5.03 | −74.2 ± 2.4 |
Data are presented as mean ± SD (HU). Original: RMSE by subtracting between input and target images. Converted: RMSE by subtracting converted images from target images. Differences: 100 × (converted RMSE - original RMSE) / original RMSE. HU = Hounsfield unit, RMSE = root mean square error, SD = standard deviation