Table 1.
Experiment Code | Number of Spectra | Type of Experiment | Sample | Sampling Type | Comments |
---|---|---|---|---|---|
Required Experiments | |||||
D1A | 30 | 1H,15N gHSQC1 | SSS2 | US3 | Field dependent acquisition:20 ms in t1 |
D2A | 39 | 1H,13C gHSQC | SSS | US | Field independent acquisition:128 total points in t1 |
D2B | 32 | 1H,13C gHSQC | SSS | NUS4 | 50% NUS of D2A |
D2C | 41 | 1H,13C gHSQC | SSS | US | Field dependent acquisition:25 ms in t1 |
D2D | 31 | 1H,13C gHSQC | SSS | NUS | 50% NUS of D2C |
D2E | 26 | 1H,13C gHSQC | SSS | NUS | Twice the scans per increment, 50% NUS of D2C |
D3A | 47 | 1H,13C gHSQC | NIST-Fab | US | Field dependent acquisition:25 ms in t1 |
D3B | 36 | 1H,13C gHSQC | NIST-Fab | NUS | 50% NUS of D3A |
Optional Experiments | |||||
E1 | 7 | 1H,13C gHSQC | NIST-Fab | Either | One parameter change from required experiments |
E1A | 45 | 1H,13C gHSQC | Either | NUS | Customized sampling schedule generated by individual laboratory |
E1B | 48 | 1H,13C gHSQC | Either | Either | Different temperatures:15 °C, 25 °C, 45 °C, or 50 °C |
E1C | 23 | 1H,13C sfHMQC5 | Either | Either | sfHMQC pulse sequence |
E2 | 11 | 1H,15N gHSQC | Either | US | One parameter change from D1A |
E2N | 2 | 1H,15N gHSQC | SSS | NUS | Customized NUS schedule generated by individual laboratory |
E2A | 14 | 1H,15N gHSQC | SSS | US | Different temperatures:15 °C, 25 °C, 45 °C, or 50 °C |
E2B | 8 | 1H,15N pHSQC | Either | US | Phase sensitive HSQCpulse sequence |
E2C | 11 | 1H,15N sfHMQC | Either | Either | SOFAST-HMQC pulse sequence |
1gHSQC = gradient selected heteronuclear single quantum coherence spectroscopy
2SSS = system suitability sample
3US = uniform sampling
4NUS = non-uniform sampling
5sfHMQC = selective optimized flip angle short transient (SOFAST) heteronuclear multiple quantum coherence spectroscopy (HMQC)