Skip to main content
. 2018 Sep 28;8(4):112–118. doi: 10.1556/1886.2018.00020

Table 2.

LAMP reaction conditions on the Genie II Mk2 system (Amplex Diagnostics GmbH)

Software on the LAMP devices eazyReport™-IVD-software (Amplex Diagnostics GmbH)
Master mix Isothermal master mix ISO-001 (OptiGene Ltd., Horsham, UK)
Dilution of individual LAMP primers for the mixing of primer mix 100 pmol/μL
Proportions of primer dilutions of the primers 5 pmol + 5 pmol + 20 pmol + 20 pmol + 10 pmol + 10 pmol
F3 + B3 + BIP + FIP + LF + LB in the primer mix
Composition of LAMP reaction mix 25 μL containing 2 μL primer mix, 8 μL sample, and 15 μL isothermal master mix
Reaction time 40 min
Normalization threshold value 70,000 arbitrary units
Amplification threshold value 10,000 arbitrary units
Minimum ratio 0.025
Time of first measurement 3 min after initiation of the reaction