Software on the LAMP devices |
eazyReport™-IVD-software (Amplex Diagnostics GmbH) |
Master mix |
Isothermal master mix ISO-001 (OptiGene Ltd., Horsham, UK) |
Dilution of individual LAMP primers for the mixing of primer mix |
100 pmol/μL |
Proportions of primer dilutions of the primers |
5 pmol + 5 pmol + 20 pmol + 20 pmol + 10 pmol + 10 pmol |
F3 + B3 + BIP + FIP + LF + LB in the primer mix |
|
Composition of LAMP reaction mix |
25 μL containing 2 μL primer mix, 8 μL sample, and 15 μL isothermal master mix |
Reaction time |
40 min |
Normalization threshold value |
70,000 arbitrary units |
Amplification threshold value |
10,000 arbitrary units |
Minimum ratio |
0.025 |
Time of first measurement |
3 min after initiation of the reaction |