Glycemic Metrics, Using the Complete Continuous Glucose Monitor Data, Comparing the 1-Week Sensor-Augmented Pump Run-in Period with the 48-h Artificial Pancreas Study [Mean (Standard Deviation)]

CGM metric SAP (n = 6) Closed loop (n = 6) P
Mean glucose (mg/dL) 132.3 (18.2) 136.1 (17.9) 0.69
SD glucose (mg/dL) 47.3 (7.8) 37.5 (8.6) 0.09
% Time
 <54 mg/dL 1.8 (1.9) 0.1 (0.2) 0.03*
 <60 mg/dL 2.9 (2.8) 0.4 (0.5) 0.03*
 <70 mg/dL 6.1 (5.7) 2.2 (2.1) 0.03*
 70–140 mg/dL 55.6 (10.0) 54.7 (14.8) ns
 70–180 mg/dL 78.8 (5.1) 83.1 (12.4) 0.31
 >180 mg/dL 15.1 (7.8) 14.7 (13.5) 0.84
 >250 mg/dL 2.6 (2.6) 0.6 (1.3) 0.13
 >300 mg/dL 0.8 (1.1) 0.0 (0.0) 0.25
*

Indicates a significant change as assessed by the signed-rank test.

CGM, continuous glucose monitor; ns, not significant; SAP, sensor-augmented pump.