Table 4.
Technical parameters for LDCT | For visual reviewa39 | For quantitative volumetryb |
---|---|---|
Peak kilovoltage | 100–140, adjusted for patient size | Same |
Tube current-time product (mAs) | Adjust with kVp to achieve CTDIvol ≤ 3.0 mGy in average-size patient:
|
Same |
Pitch | 0.83–1.43 (scanner specific) | Same |
Nominal section thickness (mm) | ≤2.5 mm, ≤1.0 mm preferred | ≤1.25 mm |
Section interval | ≤section thickness | Same |
Reconstruction filter | Manufacturer-specific; coronal & sagittal and maximum intensity projection reformations recommended | Highest resolution without edge enhancement |
Multidetector CT scanner | Typically 16 detector rows or greater | Minimum 16 detector rows; 64 or greater preferred |
LDCT, low-dose CT.
aFrom the American Association of Physicists in Medicine; see ref. 39 for parameters specific to numerous scanner models.
bScanner model and technical parameters should be the same for baseline and follow-up scans.