Recommended technical parameters for LDCT in lung cancer screening

Technical parameters for LDCT For visual reviewa39 For quantitative volumetryb
Peak kilovoltage 100–140, adjusted for patient size Same
Tube current-time product (mAs) Adjust with kVp to achieve CTDIvol ≤ 3.0 mGy in average-size patient:
  • Small patient (50–70 kg) = CTDIvol 0.25–2.8 mGy

  • Average patient (70–90 kg) = CTDIvol 0.5–4.3 mGy

  • Large patient (90–120 kg) = CTDIvol 1.0–5.6 mGy

Same
Pitch 0.83–1.43 (scanner specific) Same
Nominal section thickness (mm) ≤2.5 mm, ≤1.0 mm preferred ≤1.25 mm
Section interval ≤section thickness Same
Reconstruction filter Manufacturer-specific; coronal & sagittal and maximum intensity projection reformations recommended Highest resolution without edge enhancement
Multidetector CT scanner Typically 16 detector rows or greater Minimum 16 detector rows; 64 or greater preferred

LDCT, low-dose CT.

aFrom the American Association of Physicists in Medicine; see ref. 39 for parameters specific to numerous scanner models.

bScanner model and technical parameters should be the same for baseline and follow-up scans.