Table 2.
Optimal imaging parameters for planar dynamic and static image acquisition
Optimal imaging parameters for planar dynamic and static image acquisition | |
---|---|
Issue | Optimal parameter |
Acquisition | Image immediately after injection |
Positioning | Minimize neck–collimator distance |
Energy | 140 KeV, 20% |
Matrix acquisition | 128 × 128 |
Preset time | 120–300″ (s) |
Zoom factor | × 1.5 |
Collimator | LEGP or equivalent, depending on the manufacturer |