Table 1. Measurements of SNR in images acquired with different CTPA protocols and associated radiation dose.

Pitch values/SNR and radiation dose 70 kVp 80 kVp 100 kVp 120 kVp
0.9 2.2 3.2 0.9 2.2 3.2 0.9 2.2 3.2 0.9 2.2 3.2
Right main pulmonary artery 37.06±5.24 30.69±1.29 27.42±1.81 32.91±0.60 30.53±4.41 24.51±1.63 29.06±1.38 28.01±1.23 30.89±2.90 36.13±0.55 39.15±0.71 42.09±1.19
Left main pulmonary artery 40.22±3.94 29.89±0.30 20.87±2.03 35.27±1.10 28.35±1.30 25.63±0.79 34.38±1.46 31.31±1.23 36.86±2.51 58.83±1.07 47.54±2.32 67.30±3.90
Within right thrombus 4.22±0.13 4.16±0.02 3.28±0.16 4.84±0.30 5.02±0.30 4.92±0.37 6.52±0.27 6.44±0.33 6.37±0.59 10.55±0.71 9.10±0.65 10.07±1.41
Within left thrombus 6.02±0.50 4.99±0.43 4.83±0.49 6.48±0.26 6.49±0.42 4.82±0.42 13.02±0.51 12.36±1.27 10.61±1.04 13.27±1.27 11.73±0.73 10.89±1.09
CTDIvol (mGy) 0.41 0.15 0.12 0.42 0.21 0.21 1.01 0.55 0.55 2.22 1.05 1.05
DLP (mGy/cm) 5.90 2.70 2.20 6.20 4.00 3.60 14.60 10.30 9.50 32.20 19.70 18.10
Effective dose (mSv) 0.08 0.04 0.03 0.09 0.06 0.05 0.20 0.14 0.13 0.45 0.28 0.25

SNR, signal-to-noise ratio; CTPA, computed tomography pulmonary angiography; CTDIvol, volume CT dose index; DLP, dose length product.