Skip to main content
. 2019 Jan 30;57(2):e01629-18. doi: 10.1128/JCM.01629-18

TABLE 2.

Summary of GAS assay failures using Roche cobas Liat analyzers

Location Total no. of assays No. of failed assays Failure rate (%)
Site A
Analyzer 1 219 8 3.7
Site B
Analyzer 2 181 21 11.6
Analyzer 3 101 4 4.0
Total 501 33 6.6