Table 4.

Operational Performance of the Xpert-Point of Care System at 6 Obstetric Health Facilities, Based on Nurses Reports

Overall Xpert-PoC Test Performance N = 2736
Valid test result 2673 (97.7%)
Time between sample collection and result communication to the mother of valid test results, median (range) 110 minutes (94 minutes to 7 days)
No final valid test result 63 (2.3%)
Xpert-PoC test problem reported by nurses 183 (6.7%)
 Early termination of analysis because of power cut 106 (57.9%)
 Error/invalid result 61 (33.3%)
 Problem with computer or analyzer handling 4 (2.2%)
 Clotted blood 2 (1.1%)
 No reason indicated 10 (5.5%)
Xpert-PoC testing had to be repeated 132 (4.8%)
Xpert-PoC testing performed at other site (sample transferred) 210 (7.7%)

Abbreviation: PoC, point of care.