Table 1.

Ion exchange chromatography procedure for Ru isolation and purification.

Acid Volume (ml) Step
1° cation column: 2–4 Biorad columns filled with 10 ml of pre-cleaned AG50WX8 (200–400 mesh)
Milli-Q water 20 Resin cleaning
6M HCl 10 + 10 + 10 + 10 Resin cleaning
Milli-Q water 20 Resin cleaning
0.15 M HCl 10 Backwash
0.15 M HCl 10 Equilibration
0.15 M HCl 2.5 Load
0.15 M HCl 1 + 1 +1 + 5 Rinse HSE
6 M HCl 1 + 1 + 1 + 27 Rinse matrix
6 M HCl 10 + 10 + 10 + 10 + 10 + 10 Resin cleaning
Milli-Q water 20 Resin cleaning
2° anion column: 1 Biorad column filled with 3.8 ml of pre-cleaned AG1X8 (200–400 mesh)
Milli-Q water 10 Resin cleaning
Concentrated HNO3 10 + 10 Resin cleaning
Milli-Q water 5 Resin cleaning
Concentrated HCl 10 Resin cleaning
Milli-Q water 5 Resin cleaning
1 M HCl 3 Backwash
1 M HCl 4 Equilibration
1 M HCl 1 Load
0.5 M HCl 2 Matrix, Mo
0.8 M HNO3 0.5 Matrix, Mo
0.8 M HNO3 5 Zn, Cd, Mo
8 M HNO3 10 Ru, Re
Concentrated HNO3 4 Ru, Re
Concentrated HNO3 10 Pt, Ir
11 M HCl 2 Pt, Ir, Ag
11 M HCl 8 Pd
Concentrated HNO3 5 Pd