. 2019 Jan 18;11(1):39. doi: 10.3390/pharmaceutics11010039

Table 2.

A comparison of three protein quantification techniques; micro BCA, RP-HPLC, and HPLC-ELSD. For all three methods, the accuracy was determined at three concentrations, while Intra-day and Inter-day precision was calculated across three different concentrations with %RSD shown. The LOD and LOQ were also determined for the respective quantification techniques. Results represent the average of at least three replicate measurements ± standard deviation (SD).

Concentration R 2 Equation Mean (SD) %RSD Accuracy LOD LOQ
(µg/mL) Intra-Day Inter-Day Intra-Day Inter-Day (%) (µg/mL) (µg/mL)
micro BCA
5 0.9985 y = 0.013x + 0.0178 4.89 (0.75) 4.48 (0.73) 4.40 2.26 95.32 ± 0.94 1.85 5.61
20 20.86 (1.84) 20.61 (1.46) 2.42 1.18 102.94 ± 1.85
35 35.5 (0.79) 35.1 (0.79) 1.76 0.38 100.63 ± 0.99
RP-HPLC
25 0.9925 y = 0.3658x + 0.2811 26.02 (1.82) 27.87 (1.26) 6.41 4.35 101.39 ± 0.10 2.43 7.37
200 193.15 (18.22) 185.25 (4.24) 9.32 2.28 94.53 ± 6.03
400 436.44 (9.08) 415.40 (5.00) 2.07 1.20 102.74 ± 3.45
HPLC-ELSD
100 0.9946 y = 75,410x −2 × 106 81.17 (3.33) 87.18 (6.93) 5.89 11.44 93.37 ± 4.41 0.77 2.33
200 203.32 (5.53) 215.08 (7.90) 3.09 4.19 102.31 ± 13.8
400 417.02 (5.57) 402.31 (36.95) 1.42 9.84 99.21 ± 5.59