. 2019 Jan 9;9(1):10. doi: 10.3390/metabo9010010

Table 2.

Inter-day mean concentration, accuracy, and precision for QC standards prepared in dialyzed whole blood matrix.

[Asn]nominal (µM) [Asn]mean (µM) a Accuracy (%RE) b Precision (%CV) c
12.0 10.6 −11.8% 4.78%
200 182 −9.00% 2.31%
3200 3140 −1.89% 2.38%
[Asp]nominal (µM) [Asp]mean (µM) a Accuracy (%RE) b Precision (%CV) c
12.0 14.6 21.9% 7.05%
200 214 7.16% 3.75%
3200 3155 1.41% 2.57%
[Gln]nominal (µM) [Gln]mean (µM) a Accuracy (%RE) b Precision (%CV) c
12.0 11.1 −7.87% 5.77%
200 215 7.60% 2.32%
3200 3675 14.8% 2.50%
[Glu]nominal (µM) [Glu]mean (µM) a Accuracy (%RE) b Precision (%CV) c
12.0 14.4 20.0% 6.13%
200 209 4.36% 2.16%
3200 3122 −2.42% 1.56%

a The mean concentration was calculated from 1/x weighted linear least-squares regressions from the individual calibration curves in each batch (n = 5 over five non-consecutive days) after correcting for the endogenous amino acid content contained in the dialyzed whole blood matrix; b percent relative error: %RE = (([AA]mean/[AA]nominal)-1) * 100; c coefficient of variation.