Published in final edited form as: Invest Radiol. 2019 Mar;54(3):129–137. doi: 10.1097/RLI.0000000000000524

Table 1.

Acquisition and reconstruction parameters for the conventional energy integrating detector (EID) CT system and the prototype photon counting detector (PCD) CT system in this study. (WFBP = weighted filtered back projection; ADMIRE = advanced modeled iterative reconstruction; SAFIRE = sinogram-affirmed iterative reconstruction).

CT System EID-CT PCD-CT
Number of patients scanned 22
Scanner platform Somatom Definition AS+
Somatom Definition Flash
Somatom Definition Edge
Somatom Definition Force
Somatom Definition CounT
Collimation 128 × 0.6 mm (AS+, Flash, Edge)
192 × 0.6 mm (Force)
48 × 0.25 mm
Rotation time 0.25s to 0.5s 0.5s
Tube potential (kV) 120 140
Quality reference mAs 70 (0.5s) or 140 (0.25s) NA
Mean CTDIvol(32cm) 13.58 mGy
(25–75th percentile 10.95–16.05)
13.37 mGy (25–75th percentile 10.63–15.45)
Energy thresholds NA 25, 75 keV
Reconstruction technique WFBP (AS+, Flash and Edge)
ADMIRE (Force)
WFBP
SAFIRE
Slice thickness/increment (mm) 1.5/1.0 1.5/1.0
Reconstruction field-of-view (mm) 275 275
Reconstruction kernel/matrix size B46/512×512 (WFBP)
B46/1024×1024 (WFBP)
Bv49/512×512 (ADMIRE-Force only)
Bv49/1024×1024 (ADMIRE-Force only)
B46/1024×1024 (WFBP)
Q65/1024×1024 (SAFIRE)