Abstract
Background/Aim: Calcium phosphate cement (CPC) is used to fill bone voids in dental, orthopedic, and craniofacial applications. This study evaluated CPC marker as an injectable non-metallic fiducial marker. Materials and Methods: Six patients received 3-5 injections of CPC paste placed at a depth of 10 mm into tumors of the cervix before treatment planning CT (TPCT). Patients were treated with external-beam radiotherapy (EBRT) and high-dose rate brachytherapy (BT). We investigated marker visibility on cone-beam CT (CBCT), T2-weighted MRI, and interfraction of the marker motion for cervical cancer patients. Results: Of a total of 22 visible CPC markers at TPCT, 17 CPC markers were visible on the first CBCT. Excluding one patient, all markers were visible on CBCT during EBRT. Of 16 visible CPC markers on CBCT, 13 CPC markers were visible on the magnetic resonance imaging (MRI) obtained before BT. For CPC marker centroid movement, the mean-of-means/systematic variation/random variation were 0.2/0.4/1.4, –1.6/5.1/4.1, and –3.4/2.1/2.8 mm for the left-right, dorsal-ventral, and cranial-caudal directions, respectively. Conclusion: This is the first report of a CPC marker injected into tumors of the cervix. It can be visualized on CBCT and MRI with reductions in marker loss and artifacts.
Keywords: Calcium phosphate cement, injectable fiducial marker, cone beam computed tomography, magnetic resonance imaging, cervical cancer
As experience with image-guided radiotherapy (IGRT) grows, several studies using small gold markers (GM) for different disease sites have demonstrated increased accuracy in the delineation of the target and for patient setup (1,2).
Cone-beam computed tomography (CBCT) is an accurate and precise tool increasingly used for IGRT. This technology is composed of a kilovoltage X-ray (kV) source mounted orthogonally to the megavoltage treatment beam to acquire 3-dimensional (3D) volumetric images. Because of the poor soft-tissue contrast of CBCT, fiducial markers have been used as surrogates of the tumor position. Higher target accuracy matching has been reported using CBCT with implanted GMs than CBCT with soft tissue marking or 2D orthogonal images with implanted GMs (1). Although CBCT provides superior verification of the tumor position with implanted GMs, the higher density of GMs generates larger artifacts on CT images and compromises image quality and interpretation during on-line target localization and verification.
Calcium phosphate cement (CPC) is widely used as a bone substitute (3-5). We injected CPC paste as a fiducial marker in this study. The aim of this study was to evaluate a CPC marker as an injectable nonmetallic fiducial marker. Before starting 3D-based brachytherapy (BT) for cervical cancer in our institute, we investigated CPC marker visibility on CBCT and T2-weighted magnetic resonance imaging (MRI) for cervical cancer patients.
Materials and Methods
Patients and treatment. A prospective study was performed, including 6 cervical cancer patients diagnosed between July 2014 and January 2016. Patients were treated with external-beam radiotherapy (EBRT) to the whole pelvis with a total prescribed dose of 48.6-50.4 Gy delivered in 27-28 fractions. After 5 EBRT fractions/week were delivered at 34.2-45 Gy, two to three sessions of fractionated high-dose rate intracavitary BT were performed. The dose per fraction planned at Point A was 5-6 Gy/week. The treatment schedule included a program of 4 EBRT fractions/week combined with 1 session of BT per week. The final BT was performed within one week after the completion of EBRT. All except one patient received concurrent cisplatin at 40 mg/m2 weekly. This study was approved by the Institutional Committee (approval number: D1603029). Written informed consent for study participation was obtained from all patients.
CPC paste injections and obtained images. Biopex-R (Mitsubishi Materials Corporation, Tokyo, Japan), the injectable CPC paste, was prepared by manual mixing of 1 g of powder and 0.5 ml of liquid components for 1 min. With the patient in the lithotomy position, 0.1-0.2 ml of the paste was injected using a disposable syringe with a 19-gauge long needle using a speculum for visualization. Six patients received 3-5 injections of CPC paste placed at a depth of 10 mm into tumors of the cervix.
Treatment planning CT (TPCT) was obtained 2 h after the injection using a four-slice CT scanner (Lightspeed; GE Medical System, Waukesha, WI, USA) with custom immobilization and a 2.5-mm slice thickness. The TPCT slice dimension was 512×512 pixels with a 1×1-mm2 pixel size.
EBRT was initiated 2 days after the injection. In each fraction, the CBCT scan was acquired before dose delivery. CBCT images were acquired by an on-board CBCT system (Synergy; Elekta AB, Stockholm, Sweden) with settings of 120 kV, 25 mA, and 40 ms per frame. The image detector was shifted off-axis by 11.5 cm in the lateral direction. After the couch position was verified, CBCTs were acquired using the Elekta preset for the pelvis with 650 projections over an angle of 360˚. The reconstructed CBCT scans consisted of 410×410×264 voxels (collimated field of view (FOV)), with voxel dimensions of 1.0×1.0×1.0 mm3. Patients were positioned with reference to their skin marks, and position verification was based on registration of the bony structures in the CBCT images corresponding to TPCT images.
Visibility of the markers using T2-weighted MRI was also assessed before the first and final BT sessions. A 1.5 Tesla open MRI system (Magnetom Avanto; Siemens, Erlangen, Germany) was used to collect sequential axial images. Fifty axial images were acquired for T2-weighted turbo spin-echo sequences (TR=5970 ms, TE=102 ms, slice thickness=3 mm, 145˚flip angle, matrix 256×256, FOV 200 mm, 0.8-mm pixels).
Time schedules for obtaining the images are presented in Figure 1.
Figure 1. Time schedules of CPC paste injection, TPCT, CBCT, EBRT, and MRI. Patients with 3 sessions of BT.
Analysis. CBCT images for each fraction and two MR images were rigidly registered to the TPCT with respect to the bony anatomy, using Pinnacle3 treatment planning system version 9.2 (Philips Radiation Oncology Systems, Milpitas, CA, USA). Alignment of bony anatomies was adjusted automatically using gray-value match after manual registration. Therefore, any setup error was assumed to have been removed. Visibility of the markers using TPCT, CBCT, and T2-weighted MRI was assessed. The markers were determined visible when they were identical by two expert radiation oncologists.
To confirm the visibility of the markers, we investigated interfraction of the marker centroid motion. CPC marker analysis was performed in the X (left-right), Y (dorsal-ventral), and Z (cranial-caudal) coordinate system. To evaluate interfraction of the marker centroid motion, daily CBCT images were compared with the first CBCT image instead of the TPCT image as the CPC marker was shifted by the edematous tumor after the injections. The average position in the repeat CBCT scans, indicated by mi for patient i, was calculated as the mean of the marker centroid displacement observed in the treatment fractions. The overall mean deviation, M, was characterized by the average of all mi. The random variation (σ) was defined as the square root of the average variance. The systematic variation (Σ) was defined as the standard deviation of the averages (6).
Results
The procedure was well-tolerated. There was no interruption of the implantation due to pain or hemorrhage. There were no other procedure-related complications during or after CPC marker injection.
The clinical data and numbers of CPC markers in each of the images for the 6 patients are shown in Table I and Table II. Of the 3 CPC makers visible on TPCT in patient 5, only one marker was visible on the first CBCT, and this marker was not visible after 14 sessions of CBCT. This patient had the largest tumor volume and lowest hemoglobin level by bleeding before CPC marker injection. Of the 22 total visible CPC markers on TPCT, 5 CPC markers were not visible on the first CBCT, before the initial EBRT. Except for patient 5, all invisible CPC markers on the first CBCT were located in the center of the cervix near the cervical os on TPCT. All visible CPC markers on the first CBCT were visible on the last CBCT obtained after one or two sessions of BT.
Table I. Summary of six patients.
CPC: Calcium phosphate cement; EBRT: external-beam radiotherapy; BT: brachytherapy. *Volume measured at staging MRI, **MRI obtained before last BT.
Table II. Number of markers visible in TPCT, CBCT and MRI.
TPCT: Treatment planning CT; CBCT: cone-beam CT; BT: brachytherapy. *CBCT image obtained within 24 h before first MRI, **MRI obtained before first BT, #MRI obtained before last BT.
Of the 16 visible CPC markers on CBCT before the first BT and the last CBCT, 13 CPC markers were visible on the first and last MRI. The signal intensity of CPC was low. All invisible markers on MRI were located adjacent to the cervical canal on CBCT. Figure 2 demonstrates the visible CPC markers in 5 patients.
Figure 2. Demonstration of CPC markers in axial images (white arrows). Ureteral stents (black arrows) were inserted in patients 1 and 3 for hydronephrosis. Column 1: TPCT obtained 2 days before EBRT; column 2: CBCT obtained a day before first BT; column 3: MRI before first BT were obtained within 24 hours after the CBCT in column 2.
Patient 2 was excluded from the interfraction study of the marker motion, as the isocenter of the treatment field was located cranially and the half beam block and FOV of CBCT did not include CPC markers in the entire session. Each CPC marker was identified by the size and shape on CBCT during EBRT. With regard to the first CBCT scan, the mean-of-means/systematic variation/random variation of visible CPC markers from the second to 27th fraction of CBCT of 4 patients were 0.2/0.4/1.4, –1.6/5.1/4.1, and –3.4/2.1/2.8 mm for the left-right, dorsal-ventral, and cranial-caudal directions, respectively
Discussion
The essential criteria of fiducial markers for IGRT were visibility in the radiologic modalities involved in radiotherapeutic treatment planning and image guidance, the low production of imaging artifacts, and low perturbation of the therapeutic dose to the target volume (7).
GMs were the most widely used fiducial markers. Because the density of GMs is high, they induce markedly bright and dark streak artifacts on TPCT and CBCT images. These streak artifacts produced by GMs make treatment planning extremely difficult by obscuring the contour of soft tissue targets and reducing the accuracy in fiducial-based localization (8). The presence of the GM also affects the dose distribution near the GM surface (9,10). Besides GM, polymeric, nitinol, and carbon-coated markers are new, commercially available solid markers. These new types of solid markers were mainly developed to decrease streaking artifacts on CT (7,11).
The clinical use of liquid fiducial markers in radiotherapy was reported recently, and we have summarized them in Table III (12-19). The majority of the reports were on lipiodol, an iodized poppy-seed oil contrast medium (20), used for image guidance in the boost of bladder tumor irradiation. On 2D-kV images of the bladder, lateral projections of the bony structures of the pelvis made visualization of lipiodol very difficult, and only the anteroposterior projections showed lipiodol clearly with 0.25-0.5 ml injections (19). To be visible in the lateral projections on 2D-kV images, 0.5-2 ml of lipiodol was injected (14,15). However, to decrease the artifacts on the TPCT and CBCT scans, it was necessary to limit the volume of the marker (19,21). In addition the large artifacts on the CT images adversely affect the treatment planning system’s ability to calculate accurate dose distributions (21). Bair et al. reported the first clinical use of a hydrogel marker, polyethylene glycol hydrogel micro particles containing covalently bound iodine. This marker was used to assist with 3D-based BT in gynecologic malignancies by visualizing the target. The hydrogel material was injected at volumes ranging from 0.1 to 0.4 ml, and they concluded that injecting ≥0.4 ml of the hydrogel into the desired anatomic location is necessary to achieve unequivocal visibility on CT or MRI (12). In this study, we injected 0.1 to 0.2 ml which were the smallest injection volumes compared with the other studies and we did not evaluate 2D-kV images. Further investigation is necessary to visualize the marker in 2D-kV and investigate whether an injection volume >0.2 ml could improve visibility on MRI.
Table III. Clinical use of injectable markers in published papers.
CPC: Calcium phosphate cement; CBCT: cone beam CT; BT: brachytherapy; 2D-kV: 2-dimensional kilovoltage X-ray.
Two of the greatest advantages of using a liquid marker are that implantation can be introduced with thinner needles than solid fiducial markers and the marker size can be customized by the injection volume (21). As shown in Table III, 15–25-gauge needles were used for the lipiodol markers and 18-gauge was used for the hydrogel markers. We used a 19-gauge needle which was a smaller caliber needle than the one used for GM (12,14). Discomfort for patients and the risk of hemorrhage could be decreased using smaller needles.
Although absorbability and the ability to create 3D structures of liquid markers are considered to be advantages, there is a consideration that changes in the size and shape of markers during the course of radiation therapy may cause disadvantages for IGRT. Even the demarcations of lipiodol markers were clearly visible on CT and CBCT, they were slowly absorbed over time during the course of RT (17,19). Chai et al. stated that diffusion or deformation of a single large lipiodol spot could lead to a bias (13). Although it is difficult to conduct evaluation with a limited slice thickness and voxel dimension, we could identify each CPC marker by the size and shape on CBCT during EBRT.
Zeng et al. reported that 46 (35%) of 131 prostate patients had clearly defined calcifications in the prostate that were unambiguous even on low-dose CBCT (22). These patients had calcifications >2 mm in each direction either inside the prostate, or near the prostate border. The stability of the calcifications was investigated using the geometric center of three GMs as a reference, which is considered a surrogate of the prostate mass center. They concluded that calcifications might be a reliable natural landmark to replace seeds to localize the prostate during CBCT-guided treatment and to avoid artifacts on CT images with increasing accuracy in target delineation and dose calculation. Calcifications in the prostate were also used to evaluate the soft-tissue automatic gray-value match (23) and different similarity metrics to determine the most accurate prostate and seminal vesicle localization on daily CBCT (24).
CPC is used to fill bone voids in dental, orthopedic and craniofacial applications. Several formulations of CPC have been approved worldwide for clinical use (5). Biopex-R consists of powdered and liquid components. The powder contains 75% alphatricalcium phosphate, 18% tetracalcium phosphate, 5% dicalcium phosphate dihydrate, and 2% hydroxyapatite (HA), and the liquid contains 12% sodium succinate, 5% sodium chondroitin sulfate, and 83% distilled water. The injectable CPC paste is prepared by manual mixing of the two components for 1 min. This paste is self-hardening within 10 min and recrystallizes to HA in the living body without an organic content. They show very little bioresorbability in the living body because their final reaction product is mainly HA (4). HA is the primary mineral component of teeth and bone, being safe in the living body (3), and its enables the avoidance of artifacts on CT images. Shiotani et al. reported the implantation of CPC into soft tissue as injection laryngoplasty and revealed that the CPC was safe, non-absorbable, and effective (25).
The use of MRI has become important in treatment planning for both gynecologic and prostate cancer. Macroscopic tumor extension was visualized on MRI with a high signal intensity mass on fast or turbo spin echo sequence T2 in gynecologic cancer, and was limited when using only CT (26). T2-weighted MRI was also used as the gold standard in treatment planning for prostate cancer, because of overestimation of prostate volume with CT images compared with MRI and tumor localization was visualized on MRI (27,28). Parker et al. investigated the most suitable gold alloy marker composition for CT-MR image fusion purposes in prostate cancer. The marker localization was limited by the poor definition of the markers on MRI, and the use of fiducial markers for co-registration purposes raises the possibility of marker-induced distortion of MRI (27). Nyholm et al. suggested that internal markers appear clearly on gradient echo sequences, while there is more difficulty to identify them on T2-weighted fast or turbo spin echo sequences. The visibility of the markers was increased when the TE time was reduced, yielding a higher signal but compromising the T2-weighted contrast (29). MRI is dependent on the presence of hydrogen, but the hydrogen contents of CPC markers is very low. CPC markers are shown as black or very dark because of their low signal intensity, which is similar to that of cortical bone. Prostatic calculi are composed of calcium phosphate and organic content (30), and are usually not clearly visible on MRI. In our study, 13 CPC markers without organic content were visible as black, and 3 CPC markers adjacent to the endocervical canal were not visible, probably because of the presence of organic content.
Fiducial markers have been used in gynecological radiation therapy (11,12,31). Marked marker loss was reported by Kaatee et al. Although markers were attached to healthy parts of the cervix and attachment on the tumor was avoided, almost half of their markers were lost: half before the start of EBRT and half during EBRT. They concluded that the reason for the observed loss of markers might be tumor growth or regression, especially because it was not always possible to attach all markers to healthy cervix tissue (31). Although we did not inject CPC markers into normal tissue and placements of applicators for BT sources were performed in addition to EBRT, our marker loss was less marked compared with the results found by them. All markers were lost before or during EBRT in one patient with the lowest hemoglobin level. The reason for the marker loss might be bleeding from carcinoma of the uterine cervix.
To ensure at least a 95% dose to (on average) 99% of the CTV, Stroom et al. determined that 2 Σ + 0.7 σ is required for a margin between the CTV and PTV (6). Our CTV-to-PTV margins for the cervix calculated by their formula were 1.8, 13.1, and 6.1 mm in the left-right, dorsal-ventral, and cranial-caudal directions, respectively. Our results were lower for the left-right and cranial-caudal directions than in the report by Langerak et al. (7.1, 11.9, and 11.5 mm, respectively) and Kaatee et al. (8.9, 9.7, and 10.8 mm, respectively) (11,31). Chan et al. reported that inter-scan motion was the greatest at the fundus of the uterus, lower along the canal, and the lowest at the cervical os using serial cinematic MRI scans (32). Probably our markers were injected more central in the tumor near the cervical os, whereas others reported that markers were implanted in normal tissue away from the center of the cervical tumor.
This is the first report of non-metallic fiducial markers by injecting CPC paste into tumors of the cervix. They are visualized on CBCT and MRI while reducing marker loss and artifacts. Future studies using CPC markers for IGRT are warranted for cervical cancer and other clinical manifestations.
Conflicts of Interest
The Authors declare that there is no conflict of interest in regard to this study.
Acknowledgements
This work was supported by Japan Society for the Promotion of Science Grants-in-Aid for Scientific Research (25461922, and 17K10487).
References
- 1.Moseley DJ, White EA, Wiltshire KL, Rosewall T, Sharpe MB, Siewerdsen JH, Bissonnette JP, Gospodarowicz M, Warde P, Catton CN, Jaffray DA. Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate. Int J Radiat Oncol Biol Phys. 2007;67:942–953. doi: 10.1016/j.ijrobp.2006.10.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Soete G, De Cock M, Verellen D, Michielsen D, Keuppens F, Storme G. X-ray-assisted positioning of patients treated by conformal arc radiotherapy for prostate cancer: comparison of setup accuracy using implanted markers versus bony structures. Int J Radiat Oncol Biol Phys. 2007;67:823–827. doi: 10.1016/j.ijrobp.2006.09.041. [DOI] [PubMed] [Google Scholar]
- 3.Costantino PD, Friedman CD, Jones K, Chow LC, Sisson GA. Experimental hydroxyapatite cement cranioplasty. Plast Reconstr Surg. 1992;90:174–185. [PubMed] [Google Scholar]
- 4.Musha Y, Umeda T, Yoshizawa S, Shigemitsu T, Mizutani K, Itatani K. Effects of blood on bone cement made of calcium phosphate: problems and advantages. J Biomed Mater Res B Appl Biomater. 2010;92:95–101. doi: 10.1002/jbm.b.31493. [DOI] [PubMed] [Google Scholar]
- 5.Ambard AJ, Mueninghoff L. Calcium phosphate cement: review of mechanical and biological properties. J Prosthodont. 2006;15:321–328. doi: 10.1111/j.1532-849X.2006.00129.x. [DOI] [PubMed] [Google Scholar]
- 6.Stroom JC, de Boer HC, Huizenga H, Visser AG. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys. 1999;43:905–919. doi: 10.1016/s0360-3016(98)00468-4. [DOI] [PubMed] [Google Scholar]
- 7.Habermehl D, Henkner K, Ecker S, Jakel O, Debus J, Combs SE. Evaluation of different fiducial markers for image-guided radiotherapy and particle therapy. J Radiat Res. 2013;54:i61–68. doi: 10.1093/jrr/rrt071. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Owen R, Kron T, Foroudi F, Cox J, Zhu L, Cramb J, Sparks L, Duchesne G. The detectability and localization accuracy of implanted fiducial markers determined on in-room computerized tomography (CT) and electronic portal images (EPI) Med Dosim. 2008;33:226–233. doi: 10.1016/j.meddos.2007.12.002. [DOI] [PubMed] [Google Scholar]
- 9.Chow JC, Grigorov GN. Dose measurements near a non-radioactive gold seed using radiographic film. Phys Med Biol. 2005;50:N227–234. doi: 10.1088/0031-9155/50/18/N02. [DOI] [PubMed] [Google Scholar]
- 10.Chow JC, Grigorov GN. Monte Carlo simulations of dose near a nonradioactive gold seed. Med Phys. 2006;33:4614–4621. doi: 10.1118/1.2388573. [DOI] [PubMed] [Google Scholar]
- 11.Langerak T, Mens JW, Quint S, Bondar L, Heijkoop S, Heijmen B, Hoogeman M. Cervix motion in 50 cervical cancer patients assessed by daily cone beam computed tomographic imaging of a new type of marker. Int J Radiat Oncol Biol Phys. 2015;93:532–539. doi: 10.1016/j.ijrobp.2015.07.2261. [DOI] [PubMed] [Google Scholar]
- 12.Bair RJ, Bair E, Viswanathan AN. A radiopaque polymer hydrogel used as a fiducial marker in gynecologic-cancer patients receiving brachytherapy. Brachytherapy. 2015;14:876–880. doi: 10.1016/j.brachy.2015.08.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Chai X, van Herk M, van de Kamer JB, Remeijer P, Bex A, Betgen A, De Reijke TM, Hulshof MC, Pos FJ, Bel A. Behavior of lipiodol markers during image guided radiotherapy of bladder cancer. Int J Radiat Oncol Biol Phys. 2010;77:309–314. doi: 10.1016/j.ijrobp.2009.08.019. [DOI] [PubMed] [Google Scholar]
- 14.Dudouet P, Portalez D, Lhez JM, Elman B, Larroque JM, Bachaud JM, Redon A, Ribot JF. Trans-rectal ultrasonography (TRUS) with lipiodol injection for localization of the prostatic apex before radiotherapy planning. Radiother Oncol. 2001;61:135–141. doi: 10.1016/s0167-8140(01)00431-5. [DOI] [PubMed] [Google Scholar]
- 15.Freilich JM, Spiess PE, Biagioli MC, Fernandez DC, Shi EJ, Hunt DC, Gupta S, Wilder RB. Lipiodol as a fiducial marker for image-guided radiation therapy for bladder cancer. Int Braz J Urol. 2014;40:190–197. doi: 10.1590/S1677-5538.IBJU.2014.02.08. [DOI] [PubMed] [Google Scholar]
- 16.Meijer GJ, van der Toorn PP, Bal M, Schuring D, Weterings J, de Wildt M. High precision bladder cancer irradiation by integrating a library planning procedure of 6 prospectively generated SIB IMRT plans with image guidance using lipiodol markers. Radiother Oncol. 2012;105:174–179. doi: 10.1016/j.radonc.2012.08.011. [DOI] [PubMed] [Google Scholar]
- 17.Pos F, Bex A, Dees-Ribbers HM, Betgen A, van Herk M, Remeijer P. Lipiodol injection for target volume delineation and image guidance during radiotherapy for bladder cancer. Radiother Oncol. 2009;93:364–367. doi: 10.1016/j.radonc.2009.09.003. [DOI] [PubMed] [Google Scholar]
- 18.Rose M, Siva S, Ball D, Irving LB, Steinfort DP. Bronchoscopic delivery of lipiodol as a fiducial marker in lung tumors before radiotherapy. J Thorac Oncol. 2014;9:1579–1583. doi: 10.1097/JTO.0000000000000268. [DOI] [PubMed] [Google Scholar]
- 19.Sondergaard J, Olsen KO, Muren LP, Elstrom UV, Grau C, Hoyer M. A study of image-guided radiotherapy of bladder cancer based on lipiodol injection in the bladder wall. Acta Oncol. 2010;49:1109–1115. doi: 10.3109/02841861003789491. [DOI] [PubMed] [Google Scholar]
- 20.Bonnemain B, Guerbet M. [The history of Lipiodol (1901-1994) or How a medication may evolve with the times] Rev Hist Pharm (Paris) 1995;42:159–170. [PubMed] [Google Scholar]
- 21.Scherman Rydhog J, Irming Jolck R, Andresen TL, Munck Af Rosenschold P. Quantification and comparison of visibility and image artifacts of a new liquid fiducial marker in a lung phantom for image-guided radiation therapy. Med Phys. 2015;42:2818–2826. doi: 10.1118/1.4919616. [DOI] [PubMed] [Google Scholar]
- 22.Zeng GG, McGowan TS, Larsen TM, Bruce LM, Moran NK, Tsao JR, MacPherson MS. Calcifications are potential surrogates for prostate localization in image-guided radiotherapy. Int J Radiat Oncol Biol Phys. 2008;72:963–966. doi: 10.1016/j.ijrobp.2008.07.021. [DOI] [PubMed] [Google Scholar]
- 23.Smitsmans MH, de Bois J, Sonke JJ, Betgen A, Zijp LJ, Jaffray DA, Lebesque JV, van Herk M. Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy. Int J Radiat Oncol Biol Phys. 2005;63:975–984. doi: 10.1016/j.ijrobp.2005.07.973. [DOI] [PubMed] [Google Scholar]
- 24.Kim J, Hammoud R, Pradhan D, Zhong H, Jin RY, Movsas B, Chetty IJ. Prostate localization on daily cone-beam computed tomography images: accuracy assessment of similarity metrics. Int J Radiat Oncol Biol Phys. 2010;77:1257–1265. doi: 10.1016/j.ijrobp.2009.09.068. [DOI] [PubMed] [Google Scholar]
- 25.Shiotani A, Okubo K, Saito K, Fujimine T, Tomifuji M, Ikeda A, Araki K, Momoshima S. Injection laryngoplasty with calcium phosphate cement. Otolaryngol Head Neck Surg. 2009;140:816–821. doi: 10.1016/j.otohns.2009.01.040. [DOI] [PubMed] [Google Scholar]
- 26.Haie-Meder C, Potter R, Van Limbergen E, Briot E, De Brabandere M, Dimopoulos J, Dumas I, Hellebust TP, Kirisits C, Lang S, Muschitz S, Nevinson J, Nulens A, Petrow P, Wachter-Gerstner N, Gynaecological GECEWG Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol. 2005;74:235–245. doi: 10.1016/j.radonc.2004.12.015. [DOI] [PubMed] [Google Scholar]
- 27.Parker CC, Damyanovich A, Haycocks T, Haider M, Bayley A, Catton CN. Magnetic resonance imaging in the radiation treatment planning of localized prostate cancer using intra-prostatic fiducial markers for computed tomography co-registration. Radiother Oncol. 2003;66:217–224. doi: 10.1016/s0167-8140(02)00407-3. [DOI] [PubMed] [Google Scholar]
- 28.Hasumi M, Suzuki K, Taketomi A, Matsui H, Yamamoto T, Ito K, Kurokawa K, Aoki J, Endo K, Yamanaka H. The combination of multi-voxel MR spectroscopy with MR imaging improve the diagnostic accuracy for localization of prostate cancer. Anticancer Res. 2003;23:4223–4227. [PubMed] [Google Scholar]
- 29.Nyholm T, Nyberg M, Karlsson MG, Karlsson M. Systematisation of spatial uncertainties for comparison between a MR and a CT-based radiotherapy workflow for prostate treatments. Radiat Oncol. 2009;4:54. doi: 10.1186/1748-717X-4-54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Torres Ramirez C, Aguilar Ruiz J, Zuluaga Gomez A, Espuela Orgaz R, Del Rio Samper S. A crystallographic study of prostatic calculi. J Urol. 1980;124:840–843. doi: 10.1016/s0022-5347(17)55691-8. [DOI] [PubMed] [Google Scholar]
- 31.Kaatee RS, Olofsen MJ, Verstraate MB, Quint S, Heijmen BJ. Detection of organ movement in cervix cancer patients using a fluoroscopic electronic portal imaging device and radiopaque markers. Int J Radiat Oncol Biol Phys. 2002;54:576–583. doi: 10.1016/s0360-3016(02)02953-x. [DOI] [PubMed] [Google Scholar]
- 32.Chan P, Dinniwell R, Haider MA, Cho YB, Jaffray D, Lockwood G, Levin W, Manchul L, Fyles A, Milosevic M. Inter- and intrafractional tumor and organ movement in patients with cervical cancer undergoing radiotherapy: a cinematic-MRI point-of-interest study. Int J Radiat Oncol Biol Phys. 2008;70:1507–1515. doi: 10.1016/j.ijrobp.2007.08.055. [DOI] [PubMed] [Google Scholar]