Table 5. The analytical sensitivities and specificities for urinary total protein for 12 most common dipsticks in Croatia with pyrogallol red molybdate assay and turbidimetric method with benzethonium chloride as a references.

Pyrogallol red molybdate assay Turbidimetric method with benzethonium chloride
Dipstick Manufacturer Sensitivity Specificity Sensitivity Specificity
Combur 10 Test M Roche 69.8% 75.0% 87.2% 72.2%
ChoiceLine 10 Roche 69.2% 69.4% 81.8% 64.3%
Combur 10 Test UX Roche 66.7% 66.7% 85.7% 65.0%
ComboStik 10M DFI Co., Ltd. 60.0% 71.4% 77.8% 69.2%
ComboStik 11M DFI Co., Ltd. 60.0% 71.4% 77.8% 69.2%
CombiScreen 11SYS Analyticon 61.5% 69.4% 75.8% 66.7%
CombiScreen 10SL Analyticon 60.0% 71.4% 73.5% 68.3%
Combina 13 Human 41.0% 72.2% 55.9% 70.7%
Combina 11S Human 75.0% 45.7% 85.3% 41.5%
Combina 10M Human 70.0% 62.9% 91.7% 66.7%
UriGnost 11 BioGnost Ltd. 70.7% 70.6% 86.1% 66.7%
Multistix 10SG Siemens 67.5% 74.3% 80.0% 67.5%
Light grey fields represent the highest (≥ 80%) and dark grey fields the lowest (< 60%) sensitivities and specificities.