Table 3. Analysis of the sensitivity of bacterial, fungal and viral target detection on the Axiom Microbiome Array.
Target spiked | Genome copy number | Target detected as | Average # of probes detected /expecteda |
---|---|---|---|
Aspergillus fumigatus | 10 | Not detected | NA |
100 | A. fumigatus Af10 | 55±1 /168 | |
1000 | A. fumigatus Af10 | 81±0 /168 | |
10000 | A. fumigatus Af10 | 83±0 /168 | |
Shigella flexneri | 10 | Not detected | NA |
100 | S. flexneri 2930–71 draft (50 frags) | 2185±42 /7398 | |
1000 | S. flexneri II:(3)4 7(8) draft (196 frags) | 3781±125 /8009 | |
10000 | S. flexneri II:(3)4 7(8) draft (196 frags) | 4027±88 /8009 | |
Vaccinia lister | 100 | Not detected | NA |
1000 | Vaccinia virus strain DUKE | 78±4 /293 | |
10000 | Vaccinia virus | 148±1 /286 |
a The number of probes detected vs expected at different concentrations could be different if a different target strain was predicted by the array.