3D cone‐beam CT scanning parameters and reconstructed volumes of the various tested spine protocols on Philips hCBCT and O‐arm mCBCT for spine intraoperative imaging

hCBCT 27 cm hCBCT 37 cm hCBCT 48 cm mCBCT LD mCBCT SD mCBCT HD
Gantry size (cm) 85 70
x‐ray focal spots size (mm) 0.4 and 0.7 0.6 and 1.2
Detector size (cm) 40 × 30 40 × 30
Detector pixel matrix 2480 × 1920 2038 × 1536
Pixel size (mm) 0.154 0.194
Detector 3D digital data conversion (bits) 16 12
3D scan rotation (degrees) 180 360
Number of projections 482 302 302 391 391 750
3D scan time (sec) 8 10 10 13 13 25
Cu beam filtration (mm) 0.4 0.1
Antiscatter grid ratio 15:1 12:1
kV 120 120
mAs Variable based on estimated patient thickness by the AEC. Ranges within 50–325 64/80/160/200 for S/M/L/XL thickness 128/200/320/400 for S/M/L/XL thickness 188/300/480/600 for S/M/L/XL thickness
CBCT matrix size 2563 3833 5122 × 396 5122 × 198
CBCT voxel size (mm3) 0.4923 0.4523 0.4923 0.415 × 0.415 × 0.833
CBCT diameter‐length (cm) 12.6–12.6 17.3–17.3 25.2–19.5 21.2–16.5

LD, low dose; SD, standard dose; HD, high dose; S, small; M, medium; L, large; XL, extra large; AEC, automatic exposure control; CBCT, cone‐beam computed tomography.