Skip to main content
. 2018 Nov 22;63(23):235005. doi: 10.1088/1361-6560/aae74d

Table 1.

List of imaging parameters investigated in this study. The centroid position of the volume of interest was defined in the peak-exhale phase of the first scan and determined how all generated 2D images were positioned.

Imaging parameter Investigated settings Description
Pixel size 2.0 mm–5.0 mm in 1.0 mm steps Square pixels

Slice thickness 5.0 mm–10.0 mm in 2.5 mm steps

Imaging plane orientation −80° to 90° in 10° steps Rotation around the superior–inferior axis, rotation angle relative to the coronal plane

Acquisition strategy Slice A single plane intersecting with the centroid of the volume of interest
Stack Three parallel, adjacent imaging planes with the central one passing through the tumor or organ centroid
Cross Two orthogonal planes, intersecting at the tumor or organ centroid

Template generation strategy 2Dsingle A single template per imaging plane is extracted from the peak-exhale phase of the first scan
2Dmulti 10 templates are created from all respiratory phases of the first scan
3Dsingle 2 additional templates (3 in total) are generated from adjacent imaging planes from the peak-exhale phase, potentially increasing robustness against through-plane motion
3Dmulti Similar to 3Dsingle, 20 additional templates (30 in total) are extracted from the adjacent slices at all 10 respiratory phases