Performance measure We recommend that the endoscopy service has a planned programme of inspection, calibration, and maintenance of its clinical equipment according to the manufacturers’ advice and relevant national regulations
Domain Facilities and equipment
Category Process
Rationale This is a basic requirement to minimize the risk of equipment failure
Standards Minimum standard: a planned annual programme of inspection, calibration, and maintenance of its clinical equipment, with clinical equipment not meeting planned inspection and calibration requirements being withdrawn from use Target standard: none, the minimum standard is a safety and regulatory requirement
Consensus agreement 96.3%
PICO Not applicable
Concordance with other guidelines ASGE Yes Canada Yes EU Yes GRS/JAG accreditation Yes
Evidence grading Low/very low