Performance measure |
We recommend that the endoscopy service has a planned programme of inspection, calibration, and maintenance of its clinical equipment according to the manufacturers’ advice and relevant national regulations |
Domain |
Facilities and equipment |
Category |
Process |
Rationale |
This is a basic requirement to minimize the risk of equipment failure |
Standards |
Minimum standard: a planned annual programme of inspection, calibration, and maintenance of its clinical equipment, with clinical equipment not meeting planned inspection and calibration requirements being withdrawn from use Target standard: none, the minimum standard is a safety and regulatory requirement |
Consensus agreement |
96.3% |
PICO |
Not applicable |
Concordance with other guidelines |
ASGE Yes |
Canada Yes |
EU Yes |
GRS/JAG accreditation Yes |
Evidence grading |
Low/very low |