Study No. of Patients History of Pelvic Radiation (%) Overall Success Rate (%) Overall Cure Rate (%) % of Patients with 1 Adjustment % of patients with <1 adjustment Mean No. Adjustments Complications Mean Follow-up (mo)
Argus Sling (retropubic; Promedon, Cordoba, Argentina)
Romano 200618 48 0 83.3 73% 8.3 Dysuria (21%) Explant rate (10.4%) Urethral erosion (6%) Intra-op urethral perforation (6%) 7.5
Argus-T (trans-obturator; Promedon, Cordoba, Argentina)
Siracusano 201734 182 26.9 86.2 33.0 30.2 12.6 Transient pain (38.5%) Explant rate (9.3%) Urinary retention (8.8%) Infection (4.9%) Urethral erosion (0.5%) 22*
Male ReMeex System (MRS; Neomedic, Barcelona, Spain)
Sousa-Escandón 200717 51 19.6 84.3 64.7 100 86 Intra-op bladder perforation (9.8%) Perineal hematoma (5.9%) Varitensor infection (3.9%) Urethral erosion (2%) 32
Navalón-Monllor 201635 24 91.7 91.7 100 91 2.41 ± 1.52 1.12 ± 0.83 (MRS-II) Intra-op bladder perforation (4%) Varitensor infection (8%) Explant rate (8%) 40.7
ATOMS (A.M.I. Surgical, Vienna, Austria)
Friedl 201736 287 23 90 64 3* Explant rate (20%) 31*