Table 3.

Particle diameters and span of Man+Leu and mannitol emitted from a large-sized nasal cannula

DFR
(L/min)
D10
(μm)
D50
(μm)
D90
(μm)
Span
Man+Leu
20 1.31 ± 0.04 3.23 ± 0.24 10.6 ± 2.11 2.86 ± 0.44
25 1.23 ± 0.03 3.28 ± 0.17 12.7 ± 2.96 3.48 ± 0.83
30 1.26 ± 0.08 3.85 ± 0.14 14.8 ± 1.48 3.51 ± 0.28
Mannitol
20 1.95 ± 0.13 7.17 ± 1.43 62.1 ± 21.3 8.67 ± 3.29
25 1.69 ± 0.10 4.33 ± 0.33* 29.5 ± 7.03 6.36 ± 1.24
30 1.69 ± 0.08* 4.45 ± 0.35* 27.3 ± 5.63* 5.74 ± 1.04

The dose in all experiments was 40 ± 4 mg powder. Each value represents the mean ± SD (n = 3). D10, D50, and D90 are the particle diameters at 10, 50 and 90% of the cumulative particle size, respectively. Significant differences between DFR 20 and 25 L/min or 30 L/min are marked with an asterisk (* p < 0.05)