Table 2.

Glaucoma variables (intereye differences) at cutoff criterion for clinician and pupillometer

Clinician Pupillometer


RAPD present RAPD not present >0.3 log ≤0.3 log
CDR (cup/disk ratio) 0.14 [0.10-0.17] 0.06 [0.05-0.08]* 0.13 [0.09-0.16] 0.07 [0.05-0.08]*
HFA (MD dB) 8.56 [6.75-10.37] 3.95 [2.80-5.10]* 9.05 [7.26-10.84] 3.57 [2.49-4.64]*
OCTAve. (RNFL m) 18.60 [13.87-23.33] 9.89 [7.64-12.14]* 19.71 [15.33-24.10] 9.10 [6.88-11.32]*

Mann-Whitney U-test for all comparisons. *95% CIs is reported in brackets. RAPD test (present/not present) performed by the clinician (left) and by the pupillometer based on a 0.3 log units cut-off criterion (right). CDR is evaluated by fundoscopy, HFA is mean deviation of Humphrey field analyzer, OCTAve. is average of RNFL thickness measured via optical coherent tomography. All values represent delta between eyes