Skip to main content
. 2017 Jul 28;44(9):4928–4942. doi: 10.1002/mp.12419

Table 6.

Results of the organ dose calculations for ten patients who underwent a fluoroscopically guided procedure on the Siemens Artis zee

Organ doses (mGy)
RDSR Ka,r (mGy) Lung Pericardium Bladder Stomach Pancreas Liver Spleen Kidney (L) Kidney (R) Adrenal (L) Adrenal (R) Esophagus Gallbladder Intestines Thyroid
1106 8,476 193 107 0 40 17 34 130 35 19 62 34 121 14 4 28
1193 8,407 10 6 38 62 176 30 80 445 478 74 104 9 55 102 1
1124 8,192 38 33 22 307 797 106 599 1,719 664 613 487 55 255 191 4
1025 7,581 15 15 98 70 191 74 89 316 799 118 257 20 170 170 1
1141 6,464 10 9 78 37 97 65 27 129 628 51 114 9 99 175 2
1325 6,376 4 4 131 16 40 6 29 69 42 21 12 3 13 130 0
1166 6,300 5 3 62 6 20 30 9 31 276 11 83 4 41 70 0
1150 5,857 2 1 17 3 11 14 3 10 157 4 28 1 25 65 0
1313 5,819 1 1 81 2 2 1 4 6 4 2 2 0 1 13 0
1322 5,273 255 142 1 67 33 279 207 52 113 133 284 171 93 12 9