Abstract
This study aimed to evaluate the feasibility and accuracy of a technique for atherosclerosis imaging using local delivery of relatively small quantities (0.04–0.4 mg/kg) of labeled-specific imaging tracers targeting ICAM-1 and unpolymerized type I collagen or negative controls in 13 rabbits with atheroma induced by balloon injury in the abdominal aorta and a 12-week high-cholesterol diet. Immediately after local infusion, in vivo intravascular ultrasonography (IVUS)-NIRF imaging was performed at different time-points over a 40-minute period. The in vivo peak NIRF signal was significantly higher in the molecular tracer-injected rabbits than in the control-injected animals (P < 0.05). Ex vivo peak NIRF signal was significantly higher in the ICAM-1 probe-injected rabbits than in controls (P = 0.04), but not in the collagen probe-injected group (P = 0.29). NIRF signal discrimination following dual-probe delivery was also shown to be feasible in a single animal and thus offers the possibility of combining several distinct biological imaging agents in future studies. This innovative imaging strategy using in vivo local delivery of low concentrations of labeled molecular tracers followed by IVUS-NIRF catheter-based imaging holds potential for detection of vulnerable human coronary artery plaques.
Introduction
Cardiovascular atherosclerotic events are the leading cause of mortality worldwide1 and arise most frequently from coronary plaque disruption, which can trigger acute myocardial infarction and sudden cardiac death2. Despite our current knowledge of inflammatory and immune processes in atherogenesis, the early detection of vulnerable plaques and the ability to predict plaque rupture in susceptible patients remain an unmet need in clinical practice3,4. Although currently available intravascular imaging modalities enable in vivo characterization of morphological parameters of atheroma plaques, they have failed to demonstrate their relevance in preventing acute cardiac events5–8.
Near-infrared fluorescence (NIRF) is an innovative high-resolution imaging technology that allows visualization of inflammatory processes at the cellular level using target-specific labeled tracers, thus enabling atherosclerosis detection at early stages of the disease9–11. The concept of using specific site-targeted contrast agents for in vivo detection of vascular pathology was previously demonstrated in a canine arterial thrombi model using ultrasonic contrast agent12,13. In the past decade, NIR-fluorescence catheters have been developed for intravascular imaging that could enable in vivo detection of high-risk coronary plaque biomarkers and, consequently, optimize patient stratification and clinical management14–20. The current limitations to the clinical translation of NIRF imaging include the lack of clinically approved tracers that target specific features of plaque biology due, in part, to the weak sensitivity to molecular imaging agents, risk of toxicity from systemic administration, costs, labeling and regulatory issues21. Indocyanine green (ICG) and methylene blue are near-infrared fluorophore approved by the U.S. Food and Drug Administration (FDA) and have been used for NIRF imaging in both preclinical and clinical studies. Although Verjans et al.22 demonstrated ICG deposition in carotid plaque specimens from five patients using ex vivo intra-arterial dual-modal optical coherence tomography (OCT) – NIRF imaging, results showed less specific binding of the fluorophore to human plaque macrophages and lipids by comparison to previous data published from rabbit atherosclerotic plaques23. The discrepancy of ICG specificity observed between rabbits and human atherosclerotic plaque components could result from the short systemic half-life of ICG and the low concentration administered to subjects prior to endarterectomy22.
This preliminary study aimed at overcoming the present limitations of NIRF imaging and addressing the need for a clinically available intravascular molecular imaging modality to accurately and safely target both inflammation and specific atherosclerotic plaque components, an unmet need in clinical practice. Accordingly, the feasibility of a technique of local delivery of imaging agents was evaluated using two newly engineered near-infrared labeled molecular probes to image plaque composition and inflammation using a custom bimodal intravascular ultrasound (IVUS) – NIRF imaging catheter system in atherosclerotic rabbit aortas.
Results
Colocalization of NIRF signal in atherosclerotic plaques
Fifteen male New Zealand White (NZW) rabbits underwent balloon denudation in the distal abdominal aorta (40-mm length), followed by a 12-week high-cholesterol diet (0.5% cholesterol). In vivo IVUS-NIRF imaging was performed at week 12 following local delivery of labeled NIRF molecular agents targeting ICAM-1 and unpolymerized type I collagen at the injured site of the abdominal aorta (Fig. 1). In order to perform accurate local delivery of targeted molecular probes at the site of the injured abdominal aorta, conventional angiography of the distal aorta was performed to locate the aorto-iliac bifurcation for precise porous-balloon catheter positioning. Figure 2 shows an example of NIRF signal colocalization following local delivery of the collagen-binding peptide probe in the distal abdominal aorta of a diseased rabbit aorta. Figure 2a demonstrates the angiographic location of the balloon injury region (red square) in the distal part of the abdominal aorta, which was the site of local infusion of the molecular imaging agent. Following in situ delivery of the imaging tracer, the rabbit aorta was imaged using intravascular IVUS-NIRF imaging that enabled colocalization of in vivo NIRF signal with atherosclerotic plaque identified by IVUS (Fig. 2b). We observed that in vivo NIRF signal location obtained with the dual-mode imaging catheter matched accurately the signal obtained by ex vivo fluorescence reflectance imaging (FRI) imaging of the aorta (Fig. 2c,d). The fluorescence signal was stronger in lipid-rich segments of the aorta, as shown on white light image (Fig. 2e). Staining analysis from sections harboring atheroma plaques revealed that collagen-binding peptide fluorescence colocalized with collagen-rich plaques, as shown on Masson’s trichrome staining (Fig. 2f).
Ex vivo binding of anti-ICAM-1 single-domain antibody (sdAbs) and collagen-binding peptide to antigens in rabbit atherosclerotic plaque
Balloon denudation of rabbit abdominal aortas, in addition to atherogenic diet, led to the formation of lipid-rich atheroma plaques, as shown on histopathology sections (Fig. 3a). Staining analysis showed the presence of collagen (on Trichrome Masson’s staining) in inflamed plaques, as demonstrated by the presence of Oil Red O (ORO)-positive lipid plaques and neointima (on Hematoxylin and eosin staining). Prior to performing in vivo NIRF imaging in experimental atherosclerosis, we confirmed that both NIR-fluorescent molecular probes reacted with antigens in atherosclerotic aorta sections of hypercholesterolemic rabbits. Figure 3b shows focal staining of ICAM-1 localized in the intima of the aorta (merged image), by comparison to very weak signal obtained with the control sdAbs (Fig. 3d). Specific binding of the labeled nanobodies to ICAM-1 at the surface of cellular membranes rather than in the cytoplasm in tumor-necrosis factor alpha (TNF-α) activated human umbilical vein endothelial cells (HUVEC) was previously demonstrated24. Diffuse staining of collagen with the collagen-binding peptide was observed in both the intima and media of rabbit atherosclerotic aorta (Fig. 3c) and staining with the control collagen-peptide probe was negative (Fig. 3e).
Systemic clearance of anti-ICAM-1 sdAbs and collagen-binding peptide following in vivo local injection at the site of rabbit injured aorta
We assessed blood clearance of both molecular agents and their controls following in situ injection at the balloon-induced injury area of rabbit abdominal aorta. Serial heparinized blood samples were taken from the sheath catheter immediately following local delivery of the NIRF probes using a porous balloon and every 10 minutes to perform ex vivo fluorescence imaging quantification. The blood clearance kinetics of both anti-ICAM-1 sdAbs (Fig. 4a) and collagen-binding peptide probes (Fig. 4b) demonstrate a rapid systemic decline during the initial 30 minutes, which begins to stabilize at 40 minutes. There was no difference in blood kinetics between the molecular agents and their respective controls. Moreover, simultaneous dual-probe delivery did not alter the systemic clearance of the imaging agents by comparison to single-probe delivery (data not shown). Following local delivery of the imaging tracers, the first pullback performed using the IVUS-NIRF catheter detected a strong saturated NIRF signal in vivo that rapidly declined in the first 20 minutes for both probes, followed by a re-increase in signal intensity between 20 to 40 minutes, by comparison to a continuous slow decline of signal intensity of the control probes (Fig. 4c,d). The absence of binding of the control tracers to the arterial wall resulted in a continuous drop in signal intensity at every time-point that was then undetectable at 45 minutes. The specific binding of molecular tracers to atheroma plaques generated a strong NIRF signal of variable intensity that could be accurately imaged by the intravascular imaging catheter after 40 minutes.
Comparison between in vivo and ex vivo near-infrared fluorescence (NIRF) imaging of rabbit atheroma plaques using both anti-ICAM-1 sdAbs and collagen-binding peptide probes
Intravascular imaging was safely performed in all 13 atheromatous rabbits. As both ICAM-1 and unpolymerized type I collagen specific probes bind to rabbit atherosclerotic plaques, we evaluated the feasibility of local injections of these tracers to locate plaque in diseased animals as a potential novel modality for accurate and safe catheter-based NIRF detection. In order to determine the optimal timing for NIRF imaging, dual-modal intracoronary IVUS-NIRF imaging was performed serially following local delivery of molecular probes over a 40-minute period. Based on in vivo fluorescence signal stability, IVUS-NIRF imaging analysis acquired at 40 minutes revealed focal NIRF signals within the injured zone following either single (Fig. 5a–d) or dual-tracer infusion (Fig. 5e,f). Specific probe-labeling using IR800CW and Alexa-660, distinct fluorophores with different emission wavelength properties, enabled dual-probe injection and in vivo and ex vivo NIRF signal differentiation in a single diseased animal. Immediately after in vivo IVUS-NIRF imaging, fluorescence reflectance imaging of resected aortas was performed and revealed NIRF activity predominantly in lipid-rich regions visible in white light images of atheroma-bearing aortas, whereby signal intensity tended to correlate with plaque burden. Interestingly, focal and diffuse NIRF signals from non-injured lipid-rich arterial segments were observed in the collagen probe-injected atheroma rabbits. Concordance between in vivo and ex vivo NIRF signal location was evaluated by comparing NIR-fluorescence signal on intravascular imaging pullbacks and resected aortas using anatomic landmarks, whereas results are displayed with a shift between fluorescence signals due to postmortem tissue shrinkage (average of 25–30%). In vivo and ex vivo NIRF signals accurately matched in all rabbits for both labeled-specific tracers (Fig. 5a,b,e). By contrast, there were no NIRF signals detected in the control-injected groups (Fig. 5c,d,f), except in 2 collagen probe-injected diseased animals that showed NIRF signal in high-density plaque regions upstream from the injection site (thoracic and abdominal aorta). For the ICAM-1 targeted probe, the in vivo peak NIRF signal was 1,47-fold greater than the control (1.47 ± 0.22 versus 1.00 ± 0.05; P = 0.032; Fig. 5g) and ex vivo peak NIRF signal was 4.9-fold greater than the control (4.94 ± 1.73 versus 1.00 ± 0.71; P = 0.032; Fig. 5g). By contrast, in vivo peak NIRF signal for the unpolymerized type I collagen targeted peptide probe was 3.9-fold greater than the control (3.96 ± 1.07 versus 1.00 ± 0.03; P = 0.036; Fig. 5h) and ex vivo peak NIRF signal was 2.3-fold greater than the control but not statistically significant (2.31 ± 1.30 versus 1.00 ± 0.74; P = 0.286; Fig. 5h).
Colocalization of microscopic NIRF signal of ICAM-1 with endothelial cells localized in rabbit atheroma plaques
Immunostaining analysis was performed to further validate colocalization of NIRF signal of ICAM-1 on endothelial cells in atheroma plaques following local infusion of anti-ICAM-1 sdAbs imaging tracer. Fig. 6 shows strong NIRF signal with confocal microscopy in the balloon-induced injured area of aortas following anti-ICAM-1 sdAbs colocalization with mouse anti-rabbit ICAM-1 monoclonal antibody (red NIRF signal), and subsequent incubation with goat anti-mouse IgG1 cross-absorbed secondary antibody (merge image, yellow NIRF signal).
Discussion
In the present study, we demonstrated the feasibility of local delivery of relatively small quantities of NIRF targeted tracers for in vivo imaging of both morphological and biological biomarkers of atherosclerotic plaques using a fully-integrated real-time intravascular bimodal IVUS-NIRF imaging catheter. This preliminary study describes an application using in vivo molecular agents for invasive imaging of both inflammation and atheroma plaque components that could be translatable for future detection of high-risk human coronary artery plaques.
In vivo NIRF catheter signal detection was obtained following in situ delivery of molecular agents targeting either ICAM-1 or unpolymerized type I collagen in the balloon-induced injured area of the distal abdominal aorta of rabbits, as well as upstream of the injected region with the collagen-binding peptide agent. In vivo NIRF signals colocalized with atherosclerotic plaques visualized by IVUS, and signal intensity correlated with plaque burden. The use of a microporous-balloon catheter for in situ fluorescent probe injection, as well as repeated in vivo IVUS-NIRF imaging pullbacks in a single animal, demonstrated the feasibility and safety of the technique and enabled determination of NIRF molecular imaging probe kinetics for optimal in vivo catheter-based imaging. The use of a low-pressure microporous balloon-catheter system confers the advantage of both bringing imaging probes in contact with the intima, while avoiding plaque disruption or vessel trauma from its semi-compliant properties that eluded balloon over-expansion. Moreover, performing proper balloon sizing to vessel diameter by QCA prior to local infusion renders this technique suitable for translational applications in future studies. The absence of in vivo NIRF signal at the vicinity of atheroma plaques following injection of specific negative controls further validated the specificity of fluorescence signal obtained with both ICAM-1 and unpolymerized type I collagen targeted imaging tracers. These findings were then confirmed by matching the whole rabbit aorta fluorescence reflectance imaging ex vivo with in vivo NIRF signal detected by intravascular NIRF, which showed signal colocalization in all rabbits with respect to both molecular imaging agents.
In vivo imaging tracer kinetics were analysed following multiple pullbacks performed at different time points using a custom dual-imaging catheter. Differences observed between in vivo clearance kinetics of molecular tracers and their controls could be the result of both tracer’s initial local diffusion of bounded probes on plaque surface, along with systemic clearance of unbound probes, leading to an initial decrease in signal intensity. A subsequent increase in fluorescence signal observed with both target-specific tracers in the aftermath of additional binding of the probes from systemic circulation/redistribution was seen, more predominantly with anti-ICAM-1 sdAbs, reaching a state of stability between 40 to 50 minutes. It is plausible that the peptide collagen probe diffused more easily in the plaque than ICAM-1 nanobodies immediately following local delivery and thus a faster saturation of the ligands was observed. Moreover, from the ex vivo imaging of the whole rabbit aortas, we’ve observed focal and diffuse NIRF signal from non-injured lipid-rich arterial segments, mostly in the thoracic aorta, which may explain less redistribution of the peptide probe at sites of the injured lipid plaques in subsequent imaging time points. Nevertheless, pharmaco-kinetic properties of both imaging tracers enable targeting diseased arterial segments prior to achieving a steady state, which makes them valuable for future studies in atherosclerosis imaging.
Numerous fluorophore-coupled imaging probes targeting a variety of molecular biomarkers have been previously tested in atherosclerotic animal models for intravascular imaging, including matrix metalloproteinases25,26, macrophages23,26, cathepsins14–16,26–28 and oxidized low-density lipoproteins (oxLDL)29. Systemic administration of a prespecified dosage of these imaging compounds several hours prior to in vivo and/or ex vivo imaging was performed to allow fluorescent probe diffusion in the atheroma plaques. Indocyanine green was also used as a fluorescent probe in experimental animal models (10 mg/kg in rabbits) and humans (0.25–2 mg/kg) for the assessment of high-risk plaques following intravenous injection of safe doses17,18,22,23. By contrast, our study demonstrated the feasibility and reliability of specific in vivo NIRF signals following local delivery of a concentration 6–50 times less than the intravenous dosage of molecular fluorescent tracers (0.04–0.4 mg/kg), with the advantages of avoiding potential risk of adverse effects or immune reactions from systemic administration, suboptimal biodistribution and tracer diffusion in regions of interest. As shown in atherosclerotic rabbits following local injection of the collagen-peptide probe, in vivo NIRF signals were observed not only in the balloon-induced injury area but also in several lipid-rich regions of the aorta. This suggests that the use of another vector for selective in vivo probe delivery for future studies in coronary arteries could enable accurate imaging of biological biomarkers of surrounding atheroma plaques while avoiding the relative risk of vascular injury or compromising blood flow from in situ tracer delivery. The injection through a 6 F guiding catheter could be a plausible alternative however would enable a semi-selective injection of tracers rather that a local selective infusion in the region of interest. Perhaps the use of a microcatheter or thrombectomy catheter available in the catheterization laboratory would convey the injection of imaging probes at the vicinity of the plaque, thus render a more selective injection without the risk of ischemia or plaque disruption. From a translational standpoint, this technical application could certainly outweigh the limitations of conventional angiography in ascertaining the true extent of disease in the arterial wall, and for better assessment of non-flow limiting coronary stenoses harboring features of plaque instability.
A postmortem study revealed that plaque rupture was by far the most frequent cause of atherothrombosis, accounting for two-thirds of sudden cardiac deaths and more than 65% of acute myocardial infarctions30. Rupture-prone plaques are defined by a large lipid-rich necrotic core with an overlying thin fibrous cap infiltrated with macrophages and lymphocytes. The vulnerable plaque, or so-called thin-cap fibroatheroma (TCFA), is subjected to factors that can alter plaque morphology and thus lead to plaque instability and rupture30,31. Although the PROSPECT (Providing Regional Observations to Study Predictors of Events in the Coronary Tree) study identified specific morphologic predictors of lesion progression in non-culprit plaques at 3 years assessed by virtual histology (VH-IVUS)5, findings that were further supported by the VIVA (VH-IVUS in Vulnerable Atherosclerosis) and ATHEROREMO-IVUS (European Collaborative Project of Inflammation and Vascular Wall Remodeling in Atherosclerosis-Intravascular ultrasound) studies6,8, plaque morphology and composition alone do not accurately predict which non-culprit plaques will cause recurrent adverse cardiac events. Thus, routine use of invasive imaging to screen for vulnerable plaques and to predict the risk of cardiovascular complications is currently proscribed32. This reinforces the potential of imaging active biological processes to provide new insights on atheroma plaque growth, healing and biomarkers involved in plaque instability. In the present study, ICAM-1 and unpolymerized type I collagen were the selected biomarkers for in vivo molecular imaging of both inflammation and plaque morphology. With the knowledge that ICAM-1 is involved in plaque formation and progression, while collagen, a major constituent of atheroma plaque, undergoes proteolysis and is less produced by smooth muscle cells (SMC) in high-risk plaques, the quantification of both biomarkers by IVUS-NIRF imaging could enable characterization of plaque morphology and harbor the potential to discriminate between stable and unstable plaques at risk of rupture, along with other inflammatory biomarkers. We have demonstrated that dual injection of molecular probes coupled with tracers harboring different excitation and emission wavelength properties enabled accurate NIRF signal discrepancy in a single animal. Perhaps the combination of several biological fluorescent agents, along with plaque morphology assessed by catheter-based technology (IVUS or OCT), may be an attractive avenue for identifying in vivo high-risk plaques in future studies.
The current study describes a reliable technique for imaging specific features of atherosclerotic plaques using local delivery of labeled specific molecular tracers followed by intravascular IVUS-NIRF imaging. Although this proof-of-principle study using novel molecular imaging probes is promising for identifying both structural and biological properties of atherosclerotic plaques, these preliminary findings should be validated in a longitudinal study using a larger animal model that features coronary atheroma with a phenotype more similar to that observed in humans (e.g. swine). Such studies may provide new insights on plaque inflammation at early stages of the disease. Given that both ICAM-1 and unpolymerized type I collagen have not yet been established as being indicative of plaque rupture and thrombotic complications, the application of this imaging methodology using a combination of newly designed tracers targeting these biomarkers could pave the way for a better understanding of the pathophysiology of high-risk atheroma plaques and disease progression, especially the contributions of plaque inflammation and tissue (collagen) remodeling, and may allow to develop future targeted therapies (personalized medicine).
Limitations
Notwithstanding that this study demonstrates the feasibility of in vivo labeled tracer-based IVUS-NIRF intravascular molecular imaging as a method for detection of inflamed atherosclerotic plaques, there are several limitations to be mentioned. First, we were unable to conduct correlation analysis of quantified NIRF signal between in vivo and ex vivo fluorescence signal detected in rabbit aortas due to the small sample size. Second, despite in vivo and ex vivo blood clearance kinetics analyses of both molecular agents for optimal timing of in vivo catheter-based imaging, their diffusion properties from plaque surface remain unknown. Since a single concentration of all tracers was tested in the present study, it remains to be determined whether a lower dose using the same administration method would provide sufficient NIRF enhancement for in vivo catheter-based signal detection. In addition, in vivo imaging was performed at different time-points over a 40-minute period and, from the late resurgence in signal intensity detected by the imaging catheter in vivo for both probes, the optimal timing for intravascular imaging may be beyond this timeline. In vivo NIRF signal was obtained with 2 collagen negative control probes in lipid-rich regions of the aorta, which probably was a result of nonspecific binding of the tracer to plaque surface which may bear an endothelial surface more permeable than non-diseased regions. Another limitation is the large diameter of the current bimodal IVUS-NIRF imaging catheter prototype due to the rigid optic fiber, which limited catheter-to-wall distance in rather small diameter distal abdominal aortas. Nevertheless, further optimization of catheter dimensions and flexibility, in addition to the validation of this novel imaging technique in a porcine atherosclerotic model with plaque biology more similar to human fibroatheroma plaques, will be required to overcome these limitations.
Methods
Animal model and experimental protocol
The atherosclerotic plaque model was generated by performing balloon denudation in the distal abdominal aorta of 15 male New Zealand White rabbits (3 kg, aged 12–13 weeks; Charles River Laboratories, Saint-Constant, QC, Canada) at week 0. Rabbits were then placed on a cholesterol-enriched diet (0.5% cholesterol, Harlan Teklad Diets, Madison, WI) for 12 consecutive weeks (Fig. 1). The experimental protocol was approved by the animal ethics committee of the Montreal Heart Institute Research Center according to the guidelines of the Canadian Council on Animal Care.
Balloon denudation
Under continuous anesthesia (aceprozamine (1 mg/kg intramuscular (IM)), buprenorphine (0.01 mg/kg IM) and inhaled isoflurane (4% v/v, Baxter, Deerfield, IL)), a 5 French (F) sheath catheter (Cordis Corporation, Miami, FL, USAs) was used to cannulate the right carotid artery. Baseline angiography of the abdominal aorta (Siemens, Germany) was performed using a 5 F right Judkins (JR) 4.0 guiding catheter (Cordis Corporation, Miami, FL, USA) inserted into the aorta over a 0.014-inch guidewire (Abbott Vascular, Santa Clara, CA). The minimal lumen diameter (MLD) of the abdominal aorta, measured on an end-diastolic frame by quantitative coronary analysis (QCA; Syngo software, ARTIS Siemens, Germany), was used for accurate PTCA balloon sizing. Under fluoroscopic guidance, balloon injury was performed in the distal 40-mm of the abdominal aorta using a standard PTCA balloon-catheter (Boston Scientific, Marlborough, MA), corresponding to a balloon/artery ratio of 1.1 to 1.2:1, that was inflated at 10 atm for 30 seconds over three consecutive times, with 30-second intervals between inflations. After balloon denudation, the sheath catheter was removed, and the right carotid artery was ligated. Post-operative care consisted in strict monitoring of the vital signs and animal behavior. Analgesia was managed by subcutaneous administration of buprenorphine (0.02 mg/kg) and ketoprofen (2 mg/kg) immediately after the procedure and on day 1 (2 doses). Post-operative infection management consisted in a prophylactic dose of Borgal (67 mg/kg) administered subcutaneously immediately after the surgery combined to a sterile dressing applied on the surgical wound, with daily surveillance of the surgical sutures.
Anti-ICAM-1 single-domain antibody
Anti-ICAM-1 VHH single-domain antibody (sdAb) or nanobodies were generated by immunizing a male lama. A VHH library was constructed and panned as described before29. After four rounds of panning against recombinant human ICAM1 (R&D systems), ICAM-specific VHH were identified by phage-ELISA. Unique VHH sequences, identified by sequencing, were sub-cloned into pSJF2 expression vector and the VHHs were expressed in TG1 bacteria as described before (US patent: 8,623,369 B2)29. The ICAM-1 sdAbs were purified by IMAC column and size-exclusion chromatography (SEC), and their binding affinities were determined by surface plasmon resonance (SPR)29. The SEC-purified sdAbs were labeled with IRDye® 800CW (LI-COR Biosciences, MA, USA) infrared dye for NIRF imaging. An ELISA was used to confirm functional reactivity of ICAM-1 sdAb before and after labeling. Briefly, wells of a Nunc96 plate were coated with the recombinant human ICAM and blocked with Strating Block (Thermo Scientific). Serial dilution of sdAbs were added to the wells and incubated for 1 hour at room temperature. Specific bindings of sdAbs were detected by rabbit anti-His tags-HRP conjugate (Bethyl lab), the color was developed by adding KPL substrate (KPLGaithersburg, MD) and the plate was read at 450 nm33. As controls, wells with no antigen (blank) and coated by ICAM-1 but used a nonrelated VHH were included in the assay.
Collagen-binding linear hairpin (CBLH) peptide
This probe was derived from a collagen-binding hairpin peptide GEWTWDDATKTWTWTE that targets specifically unpolymerized type I collagen (US patent publication: 2016/0168210 A1). The peptide portion of this probe GEWTWDDATKTWTWTEGGC was purified using HPLC from materials obtained by solid-phase peptide synthesis (SPPS) and labeled via the Cys residue with Alexa-660-monomaleimide (Thermo Fisher Scientific, MA, USA) infrared dye for NIRF imaging. A negative control peptide ETVTFTKTADDYTSEGGGC was derived from a non-binding peptide GEWTYDDATKTFTVTE (US patent publication: 2016/0168210 A1) with the Trp (W) residue replaced by Ser followed by sequence reversal. The control peptide was also conjugated to Alexa-660-monomaleimide. Both probes were further purified using HPLC after Alexa-660 conjugation and dried to powders before reconstitution in physiological saline. Final concentrations of the peptide probes were determined by use of the dry weight of the peptide powders.
Rabbit atherosclerotic aorta sections
Cryosections from rabbit atherosclerotic aortas were used to test the reactivity of both the anti-ICAM-1 sdAb and unpolymerized type I collagen peptide probes, with their respective negative controls, by ex vivo fluorescence microscopy. Cryostat sections of 8 μm were obtained and incubated with 10 μg/ml of either probe for 2 hours at room temperature. Images were taken at 20x with a confocal microscope Zeiss LSM 7 Duo (Carl Zeiss Microscopy GmbH, Göttingen, Germany) with filter sets for Alexa-660 and IR800CW fluorescence signal detection (excitation/emission, 633/641–759 nm).
In vivo intravascular ultrasound (IVUS)-near-infrared fluorescence (NIRF) imaging catheter system
The 5 F compatible custom-made imaging catheter used in this study has been previously described34,35. The catheter combines a 400 μm optical fiber that enables 360-degree fluorescence imaging and photoacoustic excitation, and an ultrasound transducer (extracted from the commercially available IVUS 45 MHz Revolution catheter; Volcano Therapeutics, Rancho Cordova, CA) for acoustic imaging and photoacoustic signal detection. The catheter has a diameter of 1.4 mm and is connected to an electronic circuit that enables synchronized acquisitions of fluorescence and IVUS imaging modalities from rotating pullbacks at an angular speed of 30 revolutions/second. The pullback speed was set to 0.5 mm/second and the frame rate was 10 images/seconds, with a total pullback length of 100 mm. Raw data is transferred to a computer bearing a custom Matlab (The MathWorks, Inc., Natick, MA) user interface that filters the signals, reconstructs, and displays the images in real time during acquisition. This single IVUS-NIRF system enable to interchange lasers and filters according to the fluorophore properties to be imaged. Fluorescence excitation was performed using either a 730-nm laser diode for IR800CW labeled probe or a 660-nm laser diode for Alexa-660 labeled probe. We’ve measured Alexa-660 with the 800-nm channel and IR800CW with the 660-nm channel with the imaging catheter and no bleed-through was detected. Emission was detected by a photomultiplier tube (Hamamatsu Photonics, Hamamatsu City, Japan) combined to either a 776-nm longpass filter (IR800CW) or a 716 ± 20 nm filter (Alexa-660).
In vivo near-infrared fluorescence (NIRF) imaging procedure
Of the 15 animals, two died after balloon denudation and were excluded; 13 animals were used to perform in vivo imaging procedures at week 12. Under general anesthesia, a 5 F sheath catheter (Cordis Corporation, Miami, FL, USA) was used to cannulate the left carotid artery. Under fluoroscopic guidance, molecular probes were injected at the injured site of the distal abdominal aorta upstream from the aorto-iliac bifurcation using a 20-mm-long semi-compliant microporous balloon catheter (ClearWayTM RX, Atrium Medical, New Hampshire, USA)36 inserted over a 0.014-inch guidewire (Abbott Vascular, Santa Clara, CA). Administration dosage for the collagen-targeting probe was 0.07–0.12 mg per rabbit (formulated in an infusion volume of 2.5 ml), which was set according to the probe dosage of 0.25–0.5 mg/kg/probe in rat or mouse systemic biodistribution studies (US patent publication: 2016/0168210 A1). The anti-ICAM-1 sdAb dosage was 0.4 mg/kg per rabbit. Accurate porous-balloon catheter sizing (balloon/artery ratio of 1.1 to 1.2:1) was determined by the minimal lumen diameter (MLD) of the abdominal aorta measured on an end-diastolic frame by quantitative coronary analysis (QCA; Syngo software, ARTIS Siemens, Germany). The infusion was done over 90-sec, at 4 atm, followed by a flush of saline (1.5 mL). Seven rabbits had a single injection of either the probe targeting ICAM-1, unpolymerized type I collagen, or their respective negative controls at the injury site of the distal abdominal aorta; six rabbits had a dual-probe injection of either the combined molecular tracers or the combined controls. Immediately after the injection, the intravascular imaging catheter was inserted over a 0.014-inch guidewire and in vivo IVUS-NIRF imaging of the distal abdominal aorta was acquired every 10 minutes over a 40-minute period. For dual-probe injection, 2 consecutive pullbacks were performed at every time-points with different bandpass optical filters for fluorescence detection of both tracers. A total of 5 serial pullbacks were performed in the same arterial segment for each rabbit, whereas the aorto-iliac bifurcation corresponds to the proximal anatomic landmark of the 50-mm length pullback. Blood samples were taken from the sheath catheter (1 mL) prior to each pullback to perform ex vivo fluorescence imaging. At the end of the experiment, animals were sacrificed and the aorta was resected.
In vivo intravascular ultrasound (IVUS)-near-infrared fluorescence (NIRF) image analysis
A dedicated blood attenuation algorithm was applied to all in vivo pullbacks, as previously described35, to compensate the fluorescence signal amplitude with the distance between the imaging catheter and the arterial wall.
Ex vivo fluorescence imaging (FRI) and histopathology
After euthanasia and saline infusion, aortas were resected, opened and manually elongated in a petri dish coated with black wax and filled with saline to perform en face FRI at 37 °C using a commercial epi-illumination fluorescence imaging system (IVIS Lumina II, PerkinElmer, MA, USA). The following parameters were used: Ex/Em 745 ± 20/800 nm ± 20 nm for anti-ICAM-1 sdAbs and 640 ± 20/720 ± 20 nm for collagen-binding peptide probe, a field of view (FOV) of 10 cm, F/stop 1, Binning (M) 4/2, excitation time of 20 seconds. The total fluorescence values (intensity/cm2) of the region of interest (i.e. 20-mm injected area in the distal abdominal aorta) were analysed using a computer-based digitizing image system software (Lumina II Living Image 2.0). After each imaging session, the regions of interest (ROI; 15 mm in length) were embedded in Neg-50 (Thermo Fisher Scientific, MA, USA) and preserved at −80 °C for further microscopy validation of the ex vivo imaging results. The average tissue shrinkage was estimated between 25–30%, according to side branches that were used as landmarks and pullback length. For histopathology analyses, several 6-μm sections were cut, and morphometry was demonstrated with Masson’s trichrome staining, including collagen and fibrosis quantification. Von Kossa staining was performed to identify calcium in the neointima and media, and Oil red O (ORO, Sigma) to identify neutral lipids in the neointima and media.
Immunohistochemical staining
Cryosections (10 μm thickness) were obtained from fresh-frozen rabbit aortas. Fluorescence microscopy of aorta sections comprised in the region of interest was performed using a Zeiss LSM 7 Duo microscope (Carl Zeiss Microscopy GmbH, Göttingen, Germany) that visualized the distribution of anti-ICAM-1 NIRF signal (Ex/Em, 638/755 nm) and anti-ICAM-1 antibody (Ex/Em, 544/649 nm). NIRF signal colocalization was performed using a mouse anti-rabbit ICAM-1 monoclonal antibody, clone Rb2/3 (kindly provided by Pr. Miron Cybulsky, University of Toronto, Canada). Slides were fixed for 5 minutes in acetone at −20 °C and incubated overnight with primary antibody at 4 °C, followed by staining with a goat anti-mouse IgG1 cross-absorbed secondary antibody, Alexa Fluor 555 (A-21127, Invitrogen, Thermo Fisher Scientific, MA, USA).
Statistical analysis
Data are presented as mean ± SEM (standard error of the mean). Mann-Whitney test was used to compare peak plaque TBR between the molecular probe injection group and the control injection group for both in vivo and ex vivo fluorescence studies using Prism 7.0 (GraphPad Software, Sand Diego, CA). Values of P < 0.05 were considered statistically significant.
Acknowledgements
The authors thank Vanessa Durocher-Granger, Isabelle Croteau, Natasha Duquette, Marc-Antoine Gillis, Robert Clément, Karine Nadeau and André-François Couture D.M.V. for their participation in animal preparation and care. Feng Ni thanks the expert assistance of The-Minh Tu, Ping Wang and Ping Xu for the purification and characterization of collagen probes. We also thank Marie-Elaine Clavet for her assistance in histopathology analysis. This research was supported by the University of Montreal Endowed Research Chair in atherosclerosis and the Canada Research Chair in translational and personalized medicine to Dr Jean-Claude Tardif, a NSERC CRC Chair in Optical Vascular Imaging and CIHR grant to F. Lesage (273578). Feng Ni and Abedelnasser Abulrob were supported by the Biologics and Biomanufacturing Program of the National Research Council of Canada (NRCC Publication No. 53369).
Author Contributions
M.-J.B. wrote the manuscript. M.-J.B., P.L.L., E.R., F.L. and J.-C.T. designed the experiments. M.-J.B., M.A., F.M., P.G., T.A.-M., N.M., D.B. and P.-L.T. conducted the experiments. M.A. and F.L. previously developed the multimodal imaging system and catheter. F.N. designed and synthesized the unpolymerized type I collagen peptide probes. A.A. and M.A.-G. designed and synthesized the anti-ICAM-1 antibody. N.M., D.B., T.A.-M. and P.G. participated in the development of the animal model protocol and experiments. M.S.’s laboratory conducted the histopathology analysis. All authors reviewed the manuscript.
Data Availability
All data generated or analyzed during this study are included in this published article.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Mozaffarian D, et al. Circulation. 2015. Heart disease and stroke statistics – 2015 update: a report from the American heart association; pp. e29–322. [DOI] [PubMed] [Google Scholar]
- 2.Naghavi M, et al. From vulnerable plaque to vulnerable patient: a call for new definitions and risk assessment strategies: part I and part II. Circulation. 2003;108:1664–1672. doi: 10.1161/01.CIR.0000087480.94275.97. [DOI] [PubMed] [Google Scholar]
- 3.Ross R. Atherosclerosis: an inflammatory disease. N. Engl. J. Med. 1999;340:115–126. doi: 10.1056/NEJM199901143400207. [DOI] [PubMed] [Google Scholar]
- 4.Puri R, et al. High-risk coronary atheroma. J. Am. Coll. Cardiol. 2014;63:1134–1140. doi: 10.1016/j.jacc.2013.05.088. [DOI] [PubMed] [Google Scholar]
- 5.Stone GW, et al. A prospective natural-history study of coronary atherosclerosis. N. Engl. J. Med. 2011;364:226–235. doi: 10.1056/NEJMoa1002358. [DOI] [PubMed] [Google Scholar]
- 6.Calvert PA, et al. Association between virtual histology intravascular ultrasound findings and adverse outcomes in patients with coronary artery disease: the VIVA (VH-IVUS in Vulnerable Atherosclerosis) study. J. Am. Coll. Cardiol. Imaging. 2011;4:894–901. doi: 10.1016/j.jcmg.2011.05.005. [DOI] [PubMed] [Google Scholar]
- 7.Stone PH, et al. Prediction of progression of coronary artery disease and clinical outcomes using vascular profiling of endothelial shear stress and arterial plaque characteristics: the PREDICTION study. Circulation. 2012;126:172–181. doi: 10.1161/CIRCULATIONAHA.112.096438. [DOI] [PubMed] [Google Scholar]
- 8.Cheng JM, et al. In vivo detection of high-risk coronary plaques by radiofrequency intravascular ultrasound and cardiovascular outcome: results of the ATHEROREMO-IVUS study. Eur. Heart J. 2014;35:639–647. doi: 10.1093/eurheartj/eht484. [DOI] [PubMed] [Google Scholar]
- 9.Jaffer FA, Verjans JW. Molecular imaging of atherosclerosis: clinical state-of-the-art. Heart. 2014;100:1469–1477. doi: 10.1136/heartjnl-2011-301370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Kim JW, Jaffer FA. Emerging molecular targets for intravascular imaging of high-risk plaques. Curr. Cardiovasc. Imaging Rep. 2010;3:237–247. doi: 10.1007/s12410-010-9028-6. [DOI] [Google Scholar]
- 11.Calfon MA, Vinegoni C, Ntziachristos V, Jaffer FA. Intravascular near-infrared fluorescence molecular imaging of atherosclerosis: toward coronary arterial visualization of biologically high-risk plaques. J. Biomed. Opt. 2010;15:011107. doi: 10.1117/1.3280282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Lanza GM, et al. A novel site-targeted ultrasonic contrast agent with broad biomedical application. Circulation. 1996;94:3334–3340. doi: 10.1161/01.CIR.94.12.3334. [DOI] [PubMed] [Google Scholar]
- 13.Flacke S, et al. Novel MRI contrast agent for molecular imaging of fibrin: implications for detecting vulnerable plaques. Circulation. 2001;104:1280–1285. doi: 10.1161/hc3601.094303. [DOI] [PubMed] [Google Scholar]
- 14.Jaffer FA, et al. Real-time catheter molecular sensing of inflammation in proteolytically active atherosclerosis. Circulation. 2008;118:1802–1809. doi: 10.1161/CIRCULATIONAHA.108.785881. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Jaffer FA, et al. Two-dimensional intravascular near-infrared fluorescence molecular imaging of inflammation in atherosclerosis and stent-induced vascular injury. J. Am. Coll. Cardiol. 2011;57:2516–2526. doi: 10.1016/j.jacc.2011.02.036. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Yoo H, et al. Intra-arterial catheter for simultaneous microstructural and molecular imaging in vivo. Nat. Med. 2011;17:1680–1685. doi: 10.1038/nm.2555. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Lee S, et al. Fully integrated high-speed intravascular optical coherence tomography/near-infrared fluorescence structural/molecular imaging in vivo using a clinically available near-infrared fluorescence-emitting indocyanine green to detect inflamed lipid-rich atheromata in coronary-sized vessels. Circ. Cardiovasc. Interv. 2014;7:560–569. doi: 10.1161/CIRCINTERVENTIONS.114.001498. [DOI] [PubMed] [Google Scholar]
- 18.Kim S, et al. Intracoronary dual-modal optical coherence tomography-near-infrared fluorescence structural-molecular imaging with a clinical dose of indocyanine green for the assessment of high-risk plaques and stent-associated inflammation in a beating coronary artery. Eur. Heart J. 2016;37:2833–2844. doi: 10.1093/eurheartj/ehv726. [DOI] [PubMed] [Google Scholar]
- 19.Bozhko D, et al. Quantitative intravascular biological fluorescence-ultrasound imaging of coronary and peripheral arteries in vivo. Eur. Heart J. Cardiovasc. Imaging. 2017;18:1253–1261. doi: 10.1093/ehjci/jew222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Li Y, et al. Fully integrated optical coherence tomography, ultrasound, and indocyanine green-based fluorescence tri-modality system for intravascular imaging. Biomed Opt. Express. 2017;8:1036–1044. doi: 10.1364/BOE.8.001036. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Nutt R, Vento LJ, Ridinger MH. In vivo molecular imaging biomarkers: clinical pharmacology’s new “PET”. Clin. Pharmacol. Ther. 2007;81:792–795. doi: 10.1038/sj.clpt.6100213. [DOI] [PubMed] [Google Scholar]
- 22.Verjans JW, et al. Targeted near-infrared fluorescence imaging of atherosclerosis. J. Am. Coll. Cardiol. Imaging. 2016;9:1087–1095. doi: 10.1016/j.jcmg.2016.01.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Vinegoni C, et al. Indocyanine green enables near-infrared fluorescence imaging of lipid-rich, inflamed atherosclerotic plaques. Sci. Transl. Med. 2011;3:84ra45. doi: 10.1126/scitranslmed.3001577. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Tardif PL, et al. Validating intravascular imaging with serial optical coherence tomography and confocal fluorescence microscopy. Int. J. Mol. Sci. 2016;17:E2110. doi: 10.3390/ijms17122110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Deguchi JO, et al. Inflammation in atherosclerosis: visualizing matrix metalloproteinase action in macrophages in vivo. Circulation. 2006;114:55–62. doi: 10.1161/CIRCULATIONAHA.106.619056. [DOI] [PubMed] [Google Scholar]
- 26.Quillard T, et al. Molecular imaging of macrophage protease activity in cardiovascular inflammation in vivo. Thromb. Haemost. 2011;105:828–836. doi: 10.1160/TH10-09-0589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Chen C, et al. In vivo imaging of proteolytic activity in atherosclerosis. Circulation. 2002;105:2766–2771. doi: 10.1161/01.CIR.0000017860.20619.23. [DOI] [PubMed] [Google Scholar]
- 28.Jaffer FA, et al. Optimal visualization of cathepsin K activity in atherosclerosis with a novel, protease-activable fluorescence sensor. Circulation. 2007;115:2292–2298. doi: 10.1161/CIRCULATIONAHA.106.660340. [DOI] [PubMed] [Google Scholar]
- 29.Khamis RY, et al. Near infrared fluorescence (NIRF) molecular imaging of oxidized LDL with an autoantibody in experimental atherosclerosis. Sci Rep. 2016;6:21785. doi: 10.1038/srep21785. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Virmani R, et al. Lessons from sudden coronary death: a comprehensive morphological classification scheme for atherosclerotic lesions. Arterioscler. Thromb. Vasc. Biol. 2000;20:1262–1275. doi: 10.1161/01.ATV.20.5.1262. [DOI] [PubMed] [Google Scholar]
- 31.Muller JE, Tofler GH, Stone PH. Circadian variation and triggers of onset of acute cardiovascular disease. Circulation. 1989;79:733–743. doi: 10.1161/01.CIR.79.4.733. [DOI] [PubMed] [Google Scholar]
- 32.Mintz GS. Clinical utility of intravascular imaging and physiology in coronary artery disease. J. Am. Coll. Cardiol. 2014;64:207–222. doi: 10.1016/j.jacc.2014.01.015. [DOI] [PubMed] [Google Scholar]
- 33.Baral TN, MacKenzie R, Arbabi Ghahroudi M. Single-domain antibodies and their utilities. Curr Protoc Immunol. 103; Unit. 2013;2:17. doi: 10.1002/0471142735.im0217s103. [DOI] [PubMed] [Google Scholar]
- 34.Abran M, et al. Development of a photoacoustic, ultrasound and fluorescence imaging catheter for the study of atherosclerotic plaque. IEEE Trans. Biomed. Circuits Syst. 2014;8:696–703. doi: 10.1109/TBCAS.2014.2360560. [DOI] [PubMed] [Google Scholar]
- 35.Abran M, et al. Validating a bimodal intravascular ultrasound (IVUS) and near-infrared fluorescence (NIRF) catheter for atherosclerotic plaque detection in rabbits. Biomed. Opt. Express. 2015;6:3989–3999. doi: 10.1364/BOE.6.003989. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Saraf S, Ong PJL, Gorog DA. ClearWayTM RX – Rapid exchange therapeutic perfusion catheter. Eurointervention. 2008;3:639–642. doi: 10.4244/EIJV3I5A114. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data generated or analyzed during this study are included in this published article.