Table 2.

Analytical performance of the 7-plex immunoassay

Mean (pg/mL) Intra-assay Precision (%CV) Inter-assay Precision (%CV) LOB (pg/mL) LLOQ (pg/mL) ULOQ (pg/mL) Triplicates* (%CV)
QC1 QC2 QC1 QC2 QC1 QC2
OPN 13650.0 1435.0 4.4 6.8 5.2 14.5 278.8 389.5 40024.5 7.1
MIA 670.0 67.5 2.3 3.9 7.0 14.2 12.5 16.3 1999.0 4.7
CEACAM-1 9540.0 967.5 7.5 7.4 5.4 20.8 351.2 548.5 30079.5 5.0
MIC-1 2977.5 297.5 3.0 3.6 3.5 5.0 3.0 68.5 7407.0 3.5
SPON1 7825.0 762.5 2.9 2.9 2.8 4.5 50.8 185.3 24145.0 3.2
POSTN 8100.0 710.0 2.9 4.0 3.4 10.2 146.1 186.3 25081.3 3.4
HSP27 637.5 62.5 3.0 2.5 3.2 8.0 5.3 16.0 2001.3 2.7

QC1, high control. QC2, low control. LOB, limit of black. LLOQ, lower limit of quantitation. ULOQ, upper limit of quantification. *mean of  %CV for triplicates in all samples for each protein