Skip to main content
. 2018 May 22;8(1):50–57. doi: 10.4103/eus.eus_11_18

Table 4.

Comparison of EUS-guided fine-needle aspiration biopsy outcomes

Franseen needle (n=30), n (%) Conventional end-cut type needle (n=30), n (%) P-value
Procedure
 Trainee 27 (90) 24 (80) 0.47
 Expert 3 (10) 6 (20)
Technical success 30 (100) 30 (100) 1
Puncture route
 Transgastric 16 (53.3) 16 (53.3) 1
 Transduodenal 14 (46.7) 14 (46.7)
Number of passes
 Mean±SD 2.1±0.4 3.2±0.8 <0.001
 Range 1-3 2-5
Diagnostic accuracy 29 (96.7) 28 (93.3) 0.55
Presence of desmoplastic fibrosis 29 (96.7) 12 (40) <0.001
Presence of venous invasion 7 (23.3) 0 <0.01
Presence of lymphatic invasion 3 (10) 0 0.05
Presence of nerve invasion 1 (3.3) 0 0.33
Amount of obtained tissue (mm2)
 Mean 2.24±1.37 0.43±0.33 <0.001
 Median 2.13 0.45 <0.001
 Range 0.70-6.71 0.04-1.23
Adverse events 0 0 1

SD: Standard deviation