Skip to main content
. 2019 Feb 26;39(4):411–413. doi: 10.3343/alm.2019.39.4.411

Table 1. Analytical performance of the GC-MS assay for three phytosterols.

Within-run imprecision (%) Between-run imprecision (%) ME (%) Recovery (%) PE (%) LLOD (μmol/L) LLOQ (μmol/L)
Low High Low High Low High Low High Low High
Cholestanol 2.13 0.82 8.64 4.92 150.8 118.7 90.0 80.2 135.8 95.2 0.3 6.8
Sitosterol 4.37 2.57 7.32 9.06 119.0 107.4 94.9 83.5 112.9 89.7 0.2 12.1
Campesterol 1.64 1.16 10.94 10.23 114.8 103.3 95.5 82.2 109.6 84.9 0.3 7.5

Abbreviations: ME, matrix effect; PE, process efficiency; LLOD, lower limit of detection; LLOQ, lower limit of quantification; GC-MS, gas chromatography-mass spectrometry.