Table 4. Performance of LFIs on plasma samples using Trombley assay as bench mark assay*.

Sensitivity (95% CI) Specificity (95% CI)
SD Ebola Zaire Ag 110/156 170/171
70.51 (62.69–77.53)a 99.42 (96.78–99.99)c
ReEBOV Antigen test Kit 133/156 142/171
85.26 (78.70–90.42)b,c 83.04 (76.56–88.34)d
One step Ebola test 138/154 144/170
89.61 (83.68–93.94)b 84.71 (78.40–89.76)d
DEDIATEST EBOLA 108/154 147/171
70.13 (62.24–77.23)a 85.96 (79.84–90.80)d
altona PCR 129/156 172/172
82.69 (75.83–88.27)c 100.00 (97.88–100.00)

* Only samples with RT-PCR results for both altona and Trombley assays are included. LFI products with different superscripts for sensitivity or specificity have a statistically significant difference in performance (p<0.05)