LUCAS Versus Manual Chest Compression During Ambulance Transport: A Hemodynamic Study in a Porcine Model of Cardiac Arrest

Aurora Magliocca 1,2, Davide Olivari 1, Daria De Giorgio 1, Davide Zani 3, Martina Manfredi 3, Antonio Boccardo 3, Alberto Cucino 1,4, Giulia Sala 3, Giovanni Babini 1,4, Laura Ruggeri 1, Deborah Novelli 1, Markus B Skrifvars 5, Bjarne Madsen Hardig 6, Davide Pravettoni 3, Lidia Staszewsky 1, Roberto Latini 1, Angelo Belloli 3, Giuseppe Ristagno 1,
PMCID: PMC6405722  PMID: 30590977

Abstract

Background

Mechanical chest compression (CC) is currently suggested to deliver sustained high‐quality CC in a moving ambulance. This study compared the hemodynamic support provided by a mechanical piston device or manual CC during ambulance transport in a porcine model of cardiopulmonary resuscitation.

Methods and Results

In a simulated urban ambulance transport, 16 pigs in cardiac arrest were randomized to 18 minutes of mechanical CC with the LUCAS (n=8) or manual CC (n=8). ECG, arterial and right atrial pressure, together with end‐tidal CO2 and transthoracic impedance curve were continuously recorded. Arterial lactate was assessed during cardiopulmonary resuscitation and after resuscitation. During the initial 3 minutes of cardiopulmonary resuscitation, the ambulance was stationary, while then proceeded along a predefined itinerary. When the ambulance was stationary, CC‐generated hemodynamics were equivalent in the 2 groups. However, during ambulance transport, arterial and coronary perfusion pressure, and end‐tidal CO2 were significantly higher with mechanical CC compared with manual CC (coronary perfusion pressure: 43±4 versus 18±4 mmHg; end‐tidal CO2: 31±2 versus 19±2 mmHg, P<0.01 at 18 minutes). During cardiopulmonary resuscitation, arterial lactate was lower with mechanical CC compared with manual CC (6.6±0.4 versus 8.2±0.5 mmol/L, P<0.01). During transport, mechanical CC showed greater constancy compared with the manual CC, as represented by a higher CC fraction and a lower transthoracic impedance curve variability (P<0.01). All animals in the mechanical CC group and 6 (75%) in the manual one were successfully resuscitated.

Conclusions

This model adds evidence in favor of the use of mechanical devices to provide ongoing high‐quality CC and tissue perfusion during ambulance transport.

Keywords: ambulance transport, cardiac arrest, cardiopulmonary resuscitation, chest compression resuscitation, manual cardiopulmonary resuscitation, mechanical cardiopulmonary resuscitation

Subject Categories: Cardiopulmonary Resuscitation and Emergency Cardiac Care, Translational Studies, Cardiopulmonary Arrest


Clinical Perspective

What Is New?

  • The suggestion to use a mechanical compressor to deliver sustained high‐quality chest compression (CC) in a moving ambulance has been supported only by data on the quality of cardiopulmonary resuscitation (CPR) metrics.

  • This is the first investigation reporting greater hemodynamic support and systemic perfusion generated by mechanical CC compared with manual CC during ambulance transport in a porcine model of CPR. Mechanical CC accounted also for a better CC quality, with a lesser rescuer's physical effort requirements, compared with manual CC.

What Are the Clinical Implications?

  • The study provides evidence to the current knowledge gap on mechanical CPR during transport as claimed in the 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations.

  • The study may have a potential impact on the rescuers’ decision on whether stay on the scene or transport the cardiac arrest patient to the hospital with an ongoing mechanical CPR.

  • Indeed, the study results encourage the use of mechanical CPR devices during ambulance transport to assure ongoing high‐quality CC, adequate hemodynamic support and tissue perfusion, and rescuers’ safety.

Introduction

Out‐of‐hospital cardiac arrest is a leading cause of death worldwide.1, 2 Despite major efforts to improve outcome, the most recent trials have provided dismal end results with only 3% to 10% of patients surviving to hospital discharge.3, 4, 5, 6 Accordingly, prompt cardiopulmonary resuscitation (CPR) is the major determinant of successful resuscitation,2, 7 but its quality heterogeneity may contribute to the variable survival rates reported in different regions.8, 9

During CPR, provision of high‐quality chest compression (CC) may re‐establish systemic blood flow, achieving and maintaining threshold levels of coronary and cerebral perfusion.2, 10, 11 Nevertheless, ineffective and frequently interrupted manual CC is often provided even by well‐trained rescuers, leading to unsuccessful resuscitative efforts.12, 13, 14, 15 The challenge is even greater during transport, a condition characterized by the presence of acceleration, deceleration, and rotational forces that may affect the rescuers’ performance.16, 17, 18, 19 Thus, high‐quality manual CC in the moving ambulance is physically demanding and impractical, and might compromise providers’ safety.16, 20 For this special circumstance, the use of a mechanical CPR device, capable to deliver CC consistently, has been suggested as a reasonable alternative to manual CC.16

However, the above suggestion has been supported only by manikin studies or clinical data on the quality of CPR metrics.18, 19, 21, 22, 23, 24 Indeed, whether mechanical CPR is superior to manual CPR in special situations, such as the moving ambulance, has been highlighted as a knowledge gap in the 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science, underlying the urgency to focus research efforts on this field.16

This experimental study, therefore, sought to directly investigate the hemodynamic support generated by a mechanical piston device or manual CC in a moving ambulance. The hypothesis on whether mechanical CC would improve systemic perfusion compared with manual CC was tested in a preclinical porcine model of out‐of‐hospital cardiac arrest. The primary aim of the study was to assess if mechanical CC would provide a higher coronary perfusion pressure (CPP) compared with manual CC during ambulance transport. The secondary aim was the comparison of CC quality between the 2 CPR strategies during transport.

Methods

All procedures involving animals and their care were in conformity with national and international laws and policies (Art. 31, D. Lgs n° 26/2014). Approval of the study was obtained by the institutional review board committee and governmental institution (Ministry of Health approval no. 979/2017‐PR). The data that support the findings of this study are available from the corresponding author upon reasonable request.

Animal Preparation

Sixteen male domestic swine (34±0.5 kg) were fasted the night before the experiments except for free water access. Anesthesia was induced by intramuscular injection of ketamine (20 mg/kg) followed by intravenous administration of propofol (2 mg/kg) and sufentanyl (0.3 μg/kg) through an ear vein access. Anesthesia was maintained with a continuous intravenous infusion of propofol (4‐8 mg/kg per hour) and sufentanyl (0.3 μg/kg per hour). A cuffed tracheal tube was placed, and animals were mechanically ventilated (Bellavista 1000, IMT Medical, Switzerland) with a tidal volume of 15 mL/kg, a fraction of inspired oxygen (FiO2) of 0.21, and a positive‐end expiratory pressure of 5 cmH2O. Respiratory frequency was adjusted to maintain the end‐tidal partial pressure of carbon dioxide (EtCO2) between 35 and 40 mmHg, monitored with an infrared capnometer (LIFEPAK 15 monitor/defibrillator, Physio‐Control, WA).25

For measurement of aortic pressure, a fluid‐filled 7F catheter was advanced from the right femoral artery into the thoracic aorta. For measurements of right atrial pressure, another fluid‐filled 7F catheter was advanced from the right femoral vein into the right atrium. Conventional pressure transducers were used and connected to the monitor defibrillator (LIFEPAK 15).25 For inducing ventricular fibrillation (VF), a 5F pacing catheter was advanced from the right subclavian vein into the right ventricle.26 The position of all catheters was confirmed by characteristic pressure morphology and/or fluoroscopy. Frontal plane ECG was recorded.

Experimental Procedure

Before inducing cardiac arrest, animals were randomized by the sealed envelope method to receive either mechanical or manual CC. Animals were then placed in a standard clinical ambulance, in use at the veterinarian hospital where the experiments were performed. Baseline measurements were obtained, and VF was electrically induced with 1 to 2 mA alternating current delivered to the endocardium of the right ventricle.26 Mechanical ventilation was discontinued after onset of VF. After 2 minutes of untreated VF, continuous CC with 1 of the 2 strategies, mechanical or manual, was begun and performed for 18 minutes. Mechanical ventilation with a FiO2 of 1.0 (Oxylog, Dräger, Lübeck, Germany) was resumed simultaneously to CC (tidal volume of 500 mL, respiratory rate of 10 breaths/min).25 Every 5 minutes during CPR, epinephrine (1 mg) was administered via the right atrium, while arterial blood samples were obtained to assess lactate levels. The experimental protocol is summarized in Figure 1.

Figure 1.

Figure 1

On the top: a flowchart of the study protocol. On the bottom: a view of the ambulance cabin with an ongoing mechanical chest compression on the left and manual CC on right. CC indicates chest compression; epi, epinephrine administration; ROSC, return of spontaneous circulation; VF, ventricular fibrillation.

During the initial 3 minutes of CPR, the ambulance was stationary, and this allowed for a comparison of CC quality and hemodynamics during a basal static condition. In the following 15 minutes, the ambulance proceeded along a predefined itinerary inside the veterinarian university campus and in the surrounding area, simulating a typical urban transportation. The route of the ambulance journey, together with the total distance traveled and the average speed, were recorded using a GPS‐based tracker app (Endomondo Sports Tracker, vers. 18.6.2). Manual CC was provided in accordance to 2015 international CPR guidelines.2 A group composed by the same 4 qualified CPR providers was available for all experiments. During CPR, the rescuers could see the physiologic parameters on the defibrillator monitors, ie, arterial pressure, right atrial pressure, EtCO2, and were allowed to optimize the CPR quality accordingly, while CC rate was guided by the monitor/defibrillator metronome. Mechanical CC was delivered by the LUCAS 3.0 chest compression system (Stryker/Jolife AB, Lund, Physio‐Control, Sweden), which delivers continuous CC (rate: 102±2 per minute; depth: 53±2 mm; duty cycle: 50±5%). The mechanical compressor was already positioned on the animal chest before inducing VF.

Beside the driver and a copilot, the ambulance cabin crew consisted of 2 certified professional rescuers who alternated each other in performing manual CC every 2 minutes, and 1 operator responsible for drug administration and arterial blood sampling. A fourth investigator, seating at the head site, provided continuous timing information to the rescuers and assured compliance to the experimental protocol, without any direct intervention in the resuscitative maneuvers.

After the 18‐minute interval of CPR, defibrillation was attempted with a single biphasic 200‐joule shock, using a LIFEPAK 15 monitor/defibrillator. Return of spontaneous circulation (ROSC) was defined as the presence of sinus rhythm with a mean arterial pressure of >60 mmHg. If ROSC was not achieved, CPR was resumed and continued for 1 minute before a subsequent defibrillation with an escalating energy strategy (300‐360‐j). If VF reoccurred after ROSC, an immediate defibrillation was delivered. The same resuscitation protocol was continued until successful resuscitation or for a maximum of 5 additional minutes.

At the end of the resuscitation maneuvers, a chest computerized tomography was performed with a 16‐slices computerized tomography scanner (GE Brightspeed, GE Healthcare, Italy) to evaluate rib fractures and other major CPR‐related injuries. Animals were then returned in the operating room, where they were monitored for an additional 3 hours, under anesthesia. Catheters were then removed, wounds were repaired, and the animals were extubated and returned to their cages. Analgesia with butorphanol (0.1 mg/kg) was administered by intramuscular injection. At the end of the 72‐hour post resuscitation observation, animals were reanesthetized for echocardiographic examination and blood sample withdrawn. Animals were then euthanized painlessly with an intravenous injection of 150 mg/kg sodium thiopental.

Measurements

ECG, hemodynamics (arterial and right atrial pressures), EtCO2, and esophageal temperature were continuously recorded with 2 LIFEPAK 15 monitor/defibrillators. All data were then stored on CODE‐STAT 9.0 (Physio‐Control, WA) and subsequently exported as comma separated values (.csv) to LabChart 8.0 (ADInstruments, UK) for the analysis. The coronary perfusion pressure (CPP) was computed from the differences in time‐coincident diastolic aortic pressure and right atrial pressure.25, 26

Transthoracic echocardiography was performed using a phase‐array multifrequency 2.5‐ to 5‐MHz probe (CX50, Philips, The Netherlands). Two‐dimensional apical 4 chamber view was acquired to determine left ventricular volumes and ejection fraction calculations were computed using the modified single‐plane Simpson's rule.25 Cardiac output (CO) was determined as the product of the time‐velocity integral of the outflow curves (VTI) obtained in 5‐chamber apical view using pulsed wave Doppler, the cross sectional area of the left ventricular outflow tract (LVOT) obtained from 2‐dimensional echocardiography image in parasternal long‐axis view and heart rate (HR) [CO=VTI×LVOT×HR].

CC rate, CCs delivered per minute, and CC fraction (CCF) were calculated using the CODE‐STAT 9.0 CPR quality assessment tool, which uses the information derived from the transthoracic impedance (TTI). CC quality was additionally evaluated measuring the total power from the power spectral density analysis of the TTI curve after Fast Fourier Transformation (LabChart 8.0, ADInstrumets, UK).27 This served as a measure of variability in CC consistency (CC‐generated thoracic impedance (Impcc) variability).

CC providers’ fatigue at the end of the resuscitative maneuvers and their perception on the feasibility of CC provision during ambulance transport, intended as practicability and safety of the intervention, were evaluated using a score on a 10‐point scale from 0 (no fatigue or intervention 100% feasible) to 10 (maximal physical effort or intervention 100% impractical).

Arterial blood gases were assessed with i‐STAT System (Abbott Laboratories, Princeton, NJ). Plasma high‐sensitivity cardiac troponin T and serum neuron‐specific enolase (NSE) were measured with electrochemiluminescence assays (Roche Diagnostics, Italy).25

Functional recovery was evaluated before euthanasia according to overall performance categories as follows: 1=normal; 2=slight disability; 3=severe disability; 4=coma; and 5=brain death or death.25 Scores were assessed by veterinarian doctors masked to group treatment.

Statistical Analysis

Shapiro–Wilk test was used to confirm normal distribution of the data. Continuous variables are reported as mean±SEM or median with interquartiles [Q1–Q3], as appropriate. Categorical variables were described as count and proportion (%). For comparisons between time‐based measurements within the 2 groups, repeated‐measures analysis of variance was used. In the case of a significant test result, a post‐hoc analysis was performed using the Fisher Least Significant Difference (LSD) test. For comparisons between groups at the given time points, 1‐way analysis of variance was used. Non‐parametric Mann‐Whitney U test was used for variables not normally distributed. When the dependent variable was categorical, a Fisher exact test was used. A P<0.05 (2‐tailed) was regarded as statistically significant. GraphPad Prism 7.0 (GraphPad Software Inc., La Jolla, CA) was used for statistical analyses.

The sample size was estimated on the mean CPP. Using CPP values from a previous study,25 (38.5±13.7 mmHg after 5 minutes of mechanical CC), and assuming a 50% reduction in the manual CC group during transport compared with the mechanical CC, 8 animals per group would be needed to have a power=0.8 (α=0.05, 2‐sided).

Results

No significant differences in body weight, hemodynamics, EtCO2, cardiac function, arterial blood gases, and temperature were observed between the 2 groups at baseline (Table 1). No differences in the total distance traveled and average speed were noted between the 2 groups, as detailed in Table 2.

Table 1.

Baseline Characteristics

LUCAS (n=8) Manual (n=8)
Body weight, Kg 35±1 34±1
Heart rate, bpm 79±6 83±8
Systolic arterial pressure, mmHg 123±5 118±8
Diastolic arterial pressure, mmHg 85±6 84±4
Right atrial pressure, mmHg 5±1 5±1
End‐tidal CO2, mmHg 36±1 37±1
pH 7.44±0.02 7.44±0.01
Arterial oxygen partial pressure, mmHg 86±4 80±3
Arterial carbon dioxide partial pressure, mmHg 36±1 37±1
Arterial oxygen saturation, % 97±1 96±1
Arterial bicarbonate, mmol/L 25±2 25±1
Arterial base excess, mmol/L 1±2 1±1
Left ventricular ejection fraction, % 67±3 69±4
Left ventricular end‐diastolic volume, mL 30±3 27±2
Left ventricular end‐systolic volume, mL 10±1 8±1
hs‐cTnT, pg/mL 6 [3–8] 8 [6–9]
Temperature, °C 36.7±0.3 37.2±0.2

Data are reported in mean±SEM, except for hs‐cTnT and NSE that are expressed as median [interquartile range]. hs‐cTnT indicates highsensitivity cardiac troponin T; NSE, neuron‐specific enolase.

Table 2.

Ambulance Itinerary and Cardiopulmonary Resuscitation Quality

LUCAS CC (n=8) Manual CC (n=8)
Transport distance, km 7.8±0.4 8.6±0.5
Ambulance average speed, km/h 26.0±2 28.5±2
CC rate, n
Total CPR duration 102 [102–102] 103 [101– 104]
Static 102 [102–102]* 101 [100–101]
Transport 102 [102–102] 103 [102–105]
CC delivered per min, n
Total CPR duration 101 [100–102] 97 [93–99]
Static 97 [96–102] 100 [99–101]
Transport 102 [100–102] 97 [92–99]
CCF, %
Total CPR duration 99 [98–100] 98 [98–99]
Static 96 [95–100] 100 [100–100]
Transport 100 [99–100] 98 [97–99]
Impcc variability, ms2 2854 [1035–4584]* 16 068 [13 240–19 446]
Fatigue, score 1.2±0.3* 8.8±0.3
Feasibility, score 9.1±0.3* 3.7±0.6

Data are reported as mean±SEM or median [interquartile range]. CC indicates chest compression; CCF, chest compression fraction; CPR, cardiopulmonary resuscitation; Imp, impedance.

*P<0.01 vs manual; P<0.05.

Hemodynamics During CPR

During the initial 3 minutes of CC, performed in the static condition, CPP was equivalent in the 2 groups. However, coincident with the onset of the ambulance movement and throughout the whole transport period, CPP was significantly higher in the mechanical CC group compared with the manual one (P<0.01, Figure 2).

Figure 2.

Figure 2

Coronary perfusion pressure, end tidal CO2, and arterial lactate levels (Lac) at baseline, during cardiopulmonary resuscitation, and after return of spontaneous circulation. BL indicates baseline; CPP, coronary perfusion pressure; EtCO2, end tidal CO2; Lac, arterial lactate levels; ROSC, return of spontaneous circulation. *P<0.05, P<0.01 vs manual chest compression.

Similarly, EtCO2, systolic and diastolic arterial pressures were not different in the 2 groups during the static condition, while they were significantly higher in the mechanical CC compared with the manual one during ambulance transport (P<0.01, Figures 2 and 3). Right atrial pressure, instead, significantly increased in the manual CC group compared with the mechanical one during transport (P<0.01, Figure 3).

Figure 3.

Figure 3

Systolic (SAP) and diastolic (DAP) arterial pressure, and right atrial pressure (RAP) at baseline, during cardiopulmonary resuscitation, and after return of spontaneous circulation. BL indicates baseline; DAP, diastolic arterial pressure; RAP, right atrial pressure; ROSC, return of spontaneous circulation; SAP, systolic arterial pressure. *P<0.05, P<0.01 vs manual chest compression.

Arterial lactate showed a significantly greater increase in the manual CC group compared with mechanical one during the whole CPR period (P<0.01, Figure 2).

No differences in post‐resuscitation hemodynamics and arterial lactate were observed between the 2 groups (Figures 2 and 3).

CPR Quality and Feasibility During Transport

Data on CPR quality and feasibility are summarized in Table 2.

CC rate was similar in the 2 groups during the overall duration of CPR and complied with current guidelines recommendations.2 Nevertheless, CC rate was constant at 102 per minute in the mechanical CC group, while it slightly varied in the manual one over time (P<0.01 versus LUCAS CC). The number of CCs delivered per minute and the CCF were overall significantly higher in the mechanical CC group compared with the manual one during ambulance transport.

CC was more consistent in the mechanical CC group compared with the manual one during the whole period of CPR, as represented by a significantly lower variability in the CC‐generated TTI curve with the use of LUCAS 3 (P<0.01 versus manual CC, Table 2). More specifically, Impcc variability was similar between the 2 groups during the static condition, while it was >4‐fold greater in the manual CC group compared with the mechanical one during transport (P<0.01, Figure 4).

Figure 4.

Figure 4

LUCAS (on the left) and manual (on the right) chest compression‐generated transthoracic impedance signal (in green) and corresponding arterial pressure (in orange) during cardiopulmonary resuscitation performed in static condition (on the top) and in the moving ambulance (on the bottom). The graphs on the right represent the CC‐generated transthoracic impedance variability in the LUCAS in the manual chest compression during the static condition (on the top) and the ambulance transport (on the bottom). CC indicates chest compression. *P<0.01 vs manual chest compression.

CPR providers described the manual CC during ambulance transport as significantly more physically exhausting (P<0.01) and less feasible (P<0.01) compared with the mechanical CC (Table 2).

CPR Outcome and Survival

All 8 (100%) animals in the mechanical CC group and 6 (75%) in the manual one achieved ROSC (P=0.47, Table 3). Only a single defibrillation attempt was required before ROSC in the mechanical CC group compared with almost 2 in the manual one (P=0.06, Table 3).

Table 3.

CPR Outcome

LUCAS (n=8) Manual (n=8)
ROSC, n (%) 8/8 (100) 6/8 (75)
Defibrillations to first ROSC, n 1±0 1.8±0.4
Defibrillations to final resuscitation, n 1.8±0.4 2.5±0.6
CPR duration, min 18±0 18.7±0.4
Rib fractures, n 5±1 5±1
72‐h survival, n (%) 7/8 (88) 6/6 (100)
72‐h OPC, score 1.5±0.5 1±0
HR, bpm
PR 60 min 136±14 148±7
PR 120 min 125±14 114±8
PR 180 min 120±13 111±7
Temperature, °C
ROSC 36.9±0.2 37.2±0.2
PR 60 min 36.4±0.3 36.6±0.4
PR 120 min 36.4±0.3 36.4±0.3
PR 180 min 36.4±0.3 36.3±0.3
CO, L/min
PR 180 min 2.6±0.2 2.9±0.2
PR 72 h 4.1±0.4 3.6±0.5
EF, %
PR 180 min 60±4 53±12
PR 72 h 76±2 77±2
EDV, mL
PR 180 min 31±5 33±4
PR 72 h 38±5 37±2
ESV, mL
PR 180 min 13±4 17±6
PR 72 h 9±2 9±1
hs‐cTnT, pg/mL
PR 180 min 210 [91–619] 562 [381–687]
PR 72 h 38 [21–80] 41 [12–144]
72‐h NSE, ng/mL 0.18 [0.06–0.25] 0.16 [0.05–0.37]

Data are reported as mean±SEM, except for hs‐cTnT and NSE that are expressed as median [interquartile range]. CO indicates cardiac output; CPR, cardiopulmonary resuscitation; EDV, left ventricular end‐diastolic volume; EF, left ventricular ejection fraction; ESV, left ventricular end‐systolic volume; HR, heart rate; hs‐cTnT, high‐sensitivity cardiac troponin T; NSE, neuron‐specific enolase; OPC, overall performance category; PR, post resuscitation; ROSC, return of spontaneous circulation.

No differences in body temperature at ROSC and after resuscitation and in the total number of fractured ribs were observed between the 2 groups (Table 3).

All the resuscitated animals survived for 72 hours with a complete neurological recovery, except 1 in the mechanical CC group, which died 4 hours after resuscitation as a consequence of a hypertensive pneumothorax occurring during the transfer back to the cage (Table 3). No differences in post‐resuscitation arterial blood gases (Table 4), myocardial function, assessed by left ventricular ejection fraction and CO, and plasma levels of NSE and high‐sensitivity cardiac troponin T were observed between the 2 groups (Table 3). Nevertheless, in the early post‐resuscitation period, a consistently lower high‐sensitivity cardiac troponin T accompanied by a better ejection fraction and a lesser increased left ventricular end‐systolic volume was observed in animals subjected to mechanical CC compared with those that received manual CC (Table 3).

Table 4.

Post‐Resuscitation Arterial Blood Gas Analyses

LUCAS (n=8) Manual (n=8)
pH
PR 60 min 7.27±0.02 7.28±0.01
PR 180 min 7.40±0.01 7.41±0.02
PaO2, mmHg
PR 60 min 99±12 126±6
PR 180 min 117±11 136±3
PaCO2, mmHg
PR 60 min 44±2 44±3
PR 180 min 43±1 42±2
SpO2, %
PR 60 min 94±2 98±0
PR 180 min 98±1 99±0
HCO3, mmol/L
PR 60 min 20±1 20±1
PR 180 min 26±0 27±1
BE, mmol/L
PR 60 min −7±1 −6±1
PR 180 min 2±1 2±1

Data are reported as mean±SEM. BE indicates base excess; HCO3, bicarbonate; PaCO2; arterial carbon dioxide partial pressure; PaO2; arterial oxygen partial pressure; PR, post‐resuscitation; SpO2, arterial oxygen saturation.

Discussion

To our knowledge, this is the first investigation describing and comparing the hemodynamic support generated by a piston‐based mechanical CC versus manual CC during ambulance transport in an experimental model of cardiac arrest and CPR. This randomized, animal study demonstrated that mechanical CC allowed for a significantly greater systemic perfusion during transport, as represented by higher CPP, EtCO2, arterial pressure, and better tissue oxygenation evident with lower arterial lactate, compared with manual CC. During ambulance transport, the use of a mechanical piston compression device also accounted for better CC quality, with a lesser rescuer's physical effort requirements, compared with manual compression.

Coronary perfusion pressure is the main determinant of myocardial blood flow and threshold levels of CPP have been identified as leading predictors of CPR success.7, 10, 28, 29 Indeed, maintaining a CPP >20 mmHg has been shown to increase the likelihood of ROSC and survival in both preclinical and clinical studies.7, 29, 30, 31 In the present study, a CPP ≈ 20 mmHg was achieved in the manual CC group, but >2‐fold greater values were observed during mechanical CC. CPP generated during CPR have been shown to be directly related to the quality of CC and more specifically to the depth.10, 32 In this study, the quality of compression, derived from the TTI signal was suboptimal in the manual CC compared with the mechanical one during transport. This might have been likely associated with provision of CC with shallow depth, as previously reported in manikin studies.22, 33 Moreover, right atrial pressure significantly increased during ambulance transport in the manual group compared with the mechanical one, accounting for the lower CPP. Higher right atrial pressure in the manual CC group might have been the consequence of the suboptimal CC quality provided, which produced low CO and forward blood flow. A possible rescuers’ leaning on the animal chest to warrant a stable position against the vehicle's movements might be another valid explanation.22, 23, 34

Similarly, capnography is another valuable tool to monitor the physiological effects of CPR, as it reflects pulmonary blood flow and indirectly the CC‐generated CO.7, 10, 35 During prolonged CPR, failure to achieve an EtCO2 >10‐15 mmHg has shown a strong correlation with unsuccessful resuscitation.35, 36, 37 In this study, EtCO2 achieved the above thresholds, nevertheless, it was consistently higher in the mechanical CC group compared with the manual one during transport, anticipating a greater effectiveness of CC delivered mechanically.10, 38, 39 Somewhat surprising, during the static condition no differences in CPP, EtCO2, and hemodynamics were detected between the 2 groups, indicating a manual CC of high quality, comparable with that of the mechanical piston device. During transport, however, the use of the LUCAS provided a constant and reliable CC performance, which resulted in a higher perfusion and lesser increase in arterial lactate. The sharper increases in arterial pressure after each epinephrine administration in the mechanical CC group in contrast to the blunted response in the manual CC provides additional evidence of the better hemodynamic support generated by mechanical CC during transport.40

Adequate CC rate and CCF during CPR have been demonstrated to be associated with greater likelihood of ROSC and survival after cardiac arrest.2, 13, 14 Thus, a CC rate between 100 and 120 per minute and a CCF of at least 60% have been recommended.2 In this study, the mechanical piston device worked constantly, with a CC rate consistently stable at 102 per minute, with no variance, both in static condition and during ambulance transport. In the manual group, the CC rate fully complied with current guidelines but showed a greater variance, similarly to what has been previously reported on manikins.22, 23, 41, 42 In 4 pigs, mechanical CC needed to be interrupted immediately after onset of CPR to allow for LUCAS repositioning on the chest and this explains the unexpected lower CCF compared with the manual CC group noted during the static condition. The high CC quality, in terms of CC rate and fraction, in the manual group was likely achieved because of the presence of the metronome guide provided by the defibrillator. With this feedback, rescuers were able to compress the chest with the correct rate, even under the difficult condition created by the moving ambulance.

Deterioration of the manual CC consistency during transport has been recognized to be strongly influenced by the ambulance movements.23 Indeed, it has been reported that sudden changes in the ambulance speed may increase vibrations and induce rescuer's unnecessary movements that potentially impact on CC depth and rate, and on forces applied on the patient's chest.41 Moreover, the transport‐generated external forces, ie, acceleration, deceleration, centrifugal forces in curves, have been shown to make manual CC physically more demanding and less effective.22 In a moving ambulance, maintaining the standard 2‐handed CC technique has been also reported to be not feasible for the majority of the transport time, since providers are usually forced to perform CC with 1 hand, and the other to support themselves.43 Average speed in our study was similar in both groups as was the ambulance itinerary. However, in the animals resuscitated manually, episodes of shallow CC, leaning, altered duty cycles and compression technique, 1‐handed CC, and not correct hands position on the chest, were present. In contrast, transport seemed to have no effects on mechanical device performance, which remained stable and independent from motion influences throughout the whole ambulance journey.

Furthermore, delivery of manual CC in a moving ambulance has been described as physically exhausting, not easily reproducible, and potentially unsafe by the CPR team involved in this study. Indeed, CC performed during transport by an unrestrained provider has been considered as a hazardous situation, potentially dangerous for both the provider and the patient, and for this the mechanical devices have been suggested since they may allow providers to remain seated and restrained while CC is delivered.16, 23, 42, 44 In our study, there were no injuries, however in several instances during the transport, the rescuer's stable position was compromised and falls or nearby falls occurred, worsening the CPR quality. Moreover, additional risks for the CC provider might come from the distraction attributable to focusing on CC, and unexpected movements of the ambulance.22 Besides providing consistent high‐quality CC, mechanical compression devices might therefore significantly reduce the above described risks and improve ambulance safety practices.

This study has several strengths. The investigation provides evidence of hemodynamics generated and maintained in a moving vehicle by both a mechanical piston device and manual CC. A great effort was done to reproduce a real clinical scenario of ambulance transport with ongoing CPR in an urban area, ie, a clinical ambulance with human medical equipment and professional rescuers were used. The study results add evidence to the current knowledge gap on mechanical CPR devices as claimed in the 2015 International Consensus on CPR.16

Limitations

Some limitations deserve to be mentioned. The studies were conducted in healthy anesthetized animals and therefore in the absence of underlying diseases or injuries that are causative of cardiac arrest and with potential anesthesia‐related effects. Secondly, the time of untreated VF was relatively short, ie, 2 minutes, to be comparable with a real out‐of‐hospital cardiac arrest scenario and to account for a relevant myocardial ischemia.45 Nevertheless, the aim of the study was to investigate the hemodynamics during CPR in a moving ambulance, while effects on survival or long‐term outcome will be assessed in future studies using more clinically relevant durations of no‐flow.46 Thirdly, CC depth was not assessed, and thus the impact of transport on this CPR parameter can be only speculated based on the TTI signal and CPP. However, data on CC depth have been already reported in earlier studies performed on manikins, whereas no data on hemodynamics have been present yet. Fourthly, our rescuer team was well trained for the task requested by the experimental protocol and could optimize the CC performance based on the resulting arterial pressure and EtCO2 monitored on the defibrillators. Thus, it is likely that a CC with a quality superior than current standard has been delivered. Nevertheless, effects on hemodynamics were still not comparable with those from the LUCAS device during transport. Moreover, in accordance to what was reported in the clinical scenario,43 frequently CC providers had to perform 1‐handed CC, using the other hand to support themselves and prevent accidental falls because of transport‐generated external forces.22, 41 In these instances, the 1‐handed CC technique remains the only option to perform CC with minimal interruptions, in the absence of devices specifically designed to stabilize the provider in a moving ambulance.47 Accordingly, the efficacy of the 1‐handed CC technique compared with the standard 2‐handed approach needs future investigations. Finally, rescuers focused only on providing uninterrupted CC with no need for delivering bag ventilation because animals were mechanically ventilated. However, the use of a standardized mechanical ventilation in both groups allowed for an unbiased comparison of EtCO2 between the 2 CPR strategies.

Conclusions

In this preclinical model of CPR performed in a moving ambulance, a piston‐based mechanical CC allowed for a significantly greater hemodynamic support and systemic perfusion, as represented by higher CPP, EtCO2, and arterial pressure, and lower arterial lactate, compared with manual CC. Mechanical CC accounted also for a better CC quality, with a lesser rescuer's physical effort requirements, compared with manual compression. This study provides evidence to suggest and encourage the use of mechanical devices during ambulance transport to assure ongoing high‐quality CC, tissue perfusion, and rescuers’ safety.

Sources of Funding

The study was supported by equipment and grants from Stryker/Jolife AB, Lund, Sweden, which did not influence to any extent any of the data analysis.

Disclosures

Dr Hardig is an employee of Stryker/Jolife AB. The remaining authors have no disclosures to report.

Acknowledgments

Authors thank Dr Fredrik Arnwald Clinical Department, Stryker/Jolife AB, Lund, Sweden for having supported the proposal.

(J Am Heart Assoc. 2019;8:e011189 DOI: 10.1161/JAHA.118.011189.)

References

  • 1. Benjamin EJ, Virani SS, Callaway CW, Chamberlain AM, Chang AR, Cheng S, Chiuve SE, Cushman M, Delling FN, Deo R, de Ferranti SD, Ferguson JF, Fornage M, Gillespie C, Isasi CR, Jiménez MC, Jordan LC, Judd SE, Lackland D, Lichtman JH, Lisabeth L, Liu S, Longenecker CT, Lutsey PL, Mackey JS, Matchar DB, Matsushita K, Mussolino ME, Nasir K, O'Flaherty M, Palaniappan LP, Pandey A, Pandey DK, Reeves MJ, Ritchey MD, Rodriguez CJ, Roth GA, Rosamond WD, Sampson UKA, Satou GM, Shah SH, Spartano NL, Tirschwell DL, Tsao CW, Voeks JH, Willey JZ, Wilkins JT, Wu JH, Alger HM, Wong SS, Muntner P; American Heart Association Council on Epidemiology and Prevention Statistics Committee and Stroke Statistics Subcommittee. Heart Disease and Stroke Statistics‐2018 Update: a report from the American Heart Association. Circulation. 2018;137:e67–e492. [DOI] [PubMed] [Google Scholar]
  • 2. Perkins GD, Handley AJ, Koster RW, Castrén M, Smyth MA, Olasveengen T, Monsieurs KG, Raffay V, Gräsner JT, Wenzel V, Ristagno G, Soar J; Adult basic life support and automated external defibrillation section Collaborators . European Resuscitation Council Guidelines for Resuscitation 2015: Section 2. Adult basic life support and automated external defibrillation. Resuscitation. 2015;95:81–99. [DOI] [PubMed] [Google Scholar]
  • 3. Perkins GD, Ji C, Deakin CD, Quinn T, Nolan JP, Scomparin C, Regan S, Long J, Slowther A, Pocock H, Black JJM, Moore F, Fothergill RT, Rees N, O'Shea L, Docherty M, Gunson I, Han K, Charlton K, Finn J, Petrou S, Stallard N, Gates S, Lall R; PARAMEDIC2 Collaborators . A randomized trial of epinephrine in out‐of‐hospital cardiac arrest. N Engl J Med. 2018;379:711–721. [DOI] [PubMed] [Google Scholar]
  • 4. Benger JR, Kirby K, Black S, Brett SJ, Clout M, Lazaroo MJ, Nolan JP, Reeves BC, Robinson M, Scott LJ, Smartt H, South A, Stokes EA, Taylor J, Thomas M, Voss S, Wordsworth S, Rogers CA. Effect of a strategy of a supraglottic airway device vs tracheal intubation during out‐of‐hospital cardiac arrest on functional outcome: the AIRWAYS‐2 Randomized Clinical trial. JAMA. 2018;320:779–791. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Wang HE, Schmicker RH, Daya MR, Stephens SW, Idris AH, Carlson JN, Colella MR, Herren H, Hansen M, Richmond NJ, Puyana JCJ, Aufderheide TP, Gray RE, Gray PC, Verkest M, Owens PC, Brienza AM, Sternig KJ, May SJ, Sopko GR, Weisfeldt ML, Nichol G. Effect of a strategy of initial laryngeal tube insertion vs endotracheal intubation on 72‐hour survival in adults with out‐of‐hospital cardiac arrest: a randomized clinical trial. JAMA. 2018;320:769–778. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Soar J, Nolan JP, Böttiger BW, Perkins GD, Lott C, Carli P, Pellis T, Sandroni C, Skrifvars MB, Smith GB, Sunde K, Deakin CD; Adult advanced life support section Collaborators . European resuscitation council guidelines for resuscitation 2015 section 3. Adult advanced life support. Resuscitation. 2015;95:100–147. [DOI] [PubMed] [Google Scholar]
  • 7. Aufderheide TP, Frascone RJ, Wayne MA, Mahoney BD, Swor RA, Domeier RM, Olinger ML, Holcomb RG, Tupper DE, Yannopoulos D, Lurie KG. Standard cardiopulmonary resuscitation versus active compression‐decompression cardiopulmonary resuscitation with augmentation of negative intrathoracic pressure for out‐of‐hospital cardiac arrest: a randomised trial. Lancet. 2011;377:301–311. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Okubo M, Schmicker RH, Wallace DJ, Idris AH, Nichol G, Austin MA, Grunau B, Wittwer LK, Richmond N, Morrison LJ, Kurz MC, Cheskes S, Kudenchuk PJ, Zive DM, Aufderheide TP, Wang HE, Herren H, Vaillancourt C, Davis DP, Vilke GM, Scheuermeyer FX, Weisfeldt ML, Elmer J, Colella R, Callaway CW; Resuscitation Outcomes Consortium Investigators . Variation in survival after out‐of‐hospital cardiac arrest between emergency medical services agencies. JAMA Cardiol. 2018;3:989–999. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Gräsner JT, Lefering R, Koster RW, Masterson S, Böttiger BW, Herlitz J, Wnent J, Tjelmeland IB, Ortiz FR, Maurer H, Baubin M, Mols P, Hadžibegović I, Ioannides M, Škulec R, Wissenberg M, Salo A, Hubert H, Nikolaou NI, Lóczi G, Svavarsdóttir H, Semeraro F, Wright PJ, Clarens C, Pijls R, Cebula G, Correia VG, Cimpoesu D, Raffay V, Trenkler S, Markota A, Strömsöe A, Burkart R, Perkins GD, Bossaert LL; EuReCa ONE Collaborators . EuReCa ONE‐27 Nations, ONE Europe, ONE Registry: a prospective one month analysis of out‐of‐hospital cardiac arrest outcomes in 27 countries in Europe. Resuscitation. 2016;105:188–195. [DOI] [PubMed] [Google Scholar]
  • 10. Ristagno G, Tang W, Chang YT, Jorgenson DB, Russell JK, Huang L, Wang T, Sun S, Weil MH. The quality of chest compressions during cardiopulmonary resuscitation overrides importance of timing of defibrillation. Chest. 2007;132:70–75. [DOI] [PubMed] [Google Scholar]
  • 11. Klouche K, Weil MH, Sun S, Tang W, Povoas H, Bisera J. Stroke volumes generated by precordial compression during cardiac resuscitation. Crit Care Med. 2002;30:2626–2631. [DOI] [PubMed] [Google Scholar]
  • 12. Wik L, Kramer‐Johansen J, Myklebust H, Sørebø H, Svensson L, Fellows B, Steen PA. Quality of cardiopulmonary resuscitation during out‐of‐hospital cardiac arrest. JAMA. 2005;293:299–304. [DOI] [PubMed] [Google Scholar]
  • 13. Idris AH, Guffey D, Aufderheide TP, Brown S, Morrison LJ, Nichols P, Powell J, Daya M, Bigham BL, Atkins DL, Berg R, Davis D, Stiell I, Sopko G, Nichol G; Resuscitation Outcomes Consortium (ROC) Investigators . Relationship between chest compression rates and outcomes from cardiac arrest. Circulation. 2012;125:3004–3012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Christenson J, Andrusiek D, Everson‐Stewart S, Kudenchuk P, Hostler D, Powell J, Callaway CW, Bishop D, Vaillancourt C, Davis D, Aufderheide TP, Idris A, Stouffer JA, Stiell I, Berg R; Resuscitation Outcomes Consortium Investigators . Chest compression fraction determines survival in patients with out‐of‐hospital ventricular fibrillation. Circulation. 2009;120:1241–1247. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Perkins GD, Travers AH, Berg RA, Castren M, Considine J, Escalante R, Gazmuri RJ, Koster RW, Lim SH, Nation KJ, Olasveengen TM, Sakamoto T, Sayre MR, Sierra A, Smyth MA, Stanton D, Vaillancourt C; Basic Life Support Chapter Collaborators . Part 3: adult basic life support and automated external defibrillation 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations. Resuscitation. 2015;95:e43–e69. [DOI] [PubMed] [Google Scholar]
  • 16. Callaway CW, Soar J, Aibiki M, Böttiger BW, Brooks SC, Deakin CD, Donnino MW, Drajer S, Kloeck W, Morley PT, Morrison LJ, Neumar RW, Nicholson TC, Nolan JP, Okada K, O'Neil BJ, Paiva EF, Parr MJ, Wang TL, Witt J; Advanced Life Support Chapter Collaborators . Part 4: advanced life support: 2015 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2015;132:S84–S145. [DOI] [PubMed] [Google Scholar]
  • 17. Russi CS, Myers LA, Kolb LJ, Lohse CM, Hess EP, White RD. Comparison of chest compression quality delivered during on‐scene and ground transport cardiopulmonary resuscitation. West J Emerg Med. 2016;17:634–639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Olasveengen TM, Wik L, Steen PA. Quality of cardiopulmonary resuscitation before and during transport in out‐of‐hospital cardiac arrest. Resuscitation. 2008;76:185–190. [DOI] [PubMed] [Google Scholar]
  • 19. Sunde K, Wik L, Steen PA. Quality of mechanical, manual standard and active compression‐decompression CPR on the arrest site and during transport in a manikin model. Resuscitation. 1997;34:235–242. [DOI] [PubMed] [Google Scholar]
  • 20. Lattery DE, Silver A. The hazards of providing care in emergency vehicles: an opportunity for reform. Prehospital Emerg Care. 2009;13:388–397. [DOI] [PubMed] [Google Scholar]
  • 21. Cheskes S, Byers A, Zhan C, Verbeek PR, Ko D, Drennan IR, Buick JE, Brooks SC, Lin S, Taher A, Morrison LJ; Rescu Epistry Investigators . CPR quality during out‐of‐hospital cardiac arrest transport. Resuscitation. 2017;114:34–39. [DOI] [PubMed] [Google Scholar]
  • 22. Fox J, Fiechter R, Gerstl P, Url A, Wagner H, Lüscher TF, Eriksson U, Wyss CA. Mechanical versus manual chest compression CPR under ground ambulance transport conditions. Acute Card Care. 2013;15:1–6. [DOI] [PubMed] [Google Scholar]
  • 23. Gässler H, Ventzke MM, Lampl L, Helm M. Transport with ongoing resuscitation: a comparison between manual and mechanical compression. Emerg Med J. 2013;30:589–592. [DOI] [PubMed] [Google Scholar]
  • 24. Kim TH, Shin SD, Song KJ, Hong KJ, Ro YS, Song SW, Kim CH. Chest compression fraction between mechanical compressions on a reducible stretcher and manual compressions on a standard stretcher during transport in out‐of‐hospital cardiac arrests: the Ambulance Stretcher Innovation of Asian Cardiopulmonary Resuscitation (ASIA‐CPR) pilot trial. Prehosp Emerg Care. 2017;21:636–644. [DOI] [PubMed] [Google Scholar]
  • 25. Ristagno G, Fumagalli F, Russo I, Tantillo S, Zani DD, Locatelli V, De Maglie M, Novelli D, Staszewsky L, Vago T, Belloli A, Di Giancamillo M, Fries M, Masson S, Scanziani E, Latini R. Postresuscitation treatment with argon improves early neurological recovery in a porcine model of cardiac arrest. Shock. 2014;41:72–78. [DOI] [PubMed] [Google Scholar]
  • 26. Ristagno G, Castillo C, Tang W, Sun S, Bisera J, Weil MH. Miniaturized mechanical chest compressor: a new option for cardiopulmonary resuscitation. Resuscitation. 2008;76:191–197. [DOI] [PubMed] [Google Scholar]
  • 27. Li Y, Ristagno G, Guan J, Barbut D, Bisera J, Weil MH, Tang W. Preserved heart rate variability during therapeutic hypothermia correlated to 96 hrs neurological outcomes and survival in a pig model of cardiac arrest. Crit Care Med. 2012;40:580–586. [DOI] [PubMed] [Google Scholar]
  • 28. Halperin HR, Tsitlik JE, Guerci AD, Mellits ED, Levin HR, Shi AY, Chandra N, Weisfeldt ML. Determinants of blood flow to vital organs during cardiopulmonary resuscitation in dogs. Circulation. 1986;73:539–550. [DOI] [PubMed] [Google Scholar]
  • 29. Paradis NA, Martin GB, Rivers EP, Goetting MG, Appleton TJ, Feingold M, Nowak RM. Coronary perfusion pressure and the return of spontaneous circulation in human cardiopulmonary resuscitation. JAMA. 1990;263:1106–1113. [PubMed] [Google Scholar]
  • 30. Kern KB, Ewy GA, Voorhees WD, Babbs CF, Tacker WA. Myocardial perfusion pressure: a predictor of 24‐hour survival during prolonged cardiac arrest in dogs. Resuscitation. 1988;16:241–250. [DOI] [PubMed] [Google Scholar]
  • 31. Naim MY, Sutton RM, Friess SH, Bratinov G, Bhalala U, Kilbaugh TJ, Lampe JW, Nadkarni VM, Becker LB, Berg RA. Blood pressure‐ and coronary perfusion pressure‐targeted cardiopulmonary resuscitation improves 24‐hour survival from ventricular fibrillation cardiac arrest. Crit Care Med. 2016;44:e1111–e1117. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Wik L, Naess PA, Ilebekk A, Nicolaysen G, Steen PA. Effects of various degrees of compression and active decompression on haemodynamics, end‐tidal CO2, and ventilation during cardiopulmonary resuscitation of pigs. Resuscitation. 1996;31:45–57. [DOI] [PubMed] [Google Scholar]
  • 33. Blair L, Kendal SP, Shaw GR, Byers S, Dew R, Norton M, Wilkes S, Wright G. Comparison of manual and mechanical cardiopulmonary resuscitation on the move using a manikin: a service evaluation. British Paramedic Journal. 2017;2:6–15. [Google Scholar]
  • 34. Niles DE, Sutton RM, Nadkarni VM, Glatz A, Zuercher M, Maltese MR, Eilevstjønn J, Abella BS, Becker LB, Berg RA. Prevalence and hemodynamic effects of leaning during CPR. Resuscitation. 2011;82:S23–S26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Weil MH, Bisera J, Trevino RP, Rackow EC. Cardiac output and end‐tidal carbon dioxide. Crit Care Med. 1985;13:907–909. [DOI] [PubMed] [Google Scholar]
  • 36. Sanders AB, Kern KB, Otto CW, Milander MM, Ewy GA. End‐tidal carbon dioxide monitoring during cardiopulmonary resuscitation: a prognostic indicator for survival. JAMA. 1989;262:1347–1351. [PubMed] [Google Scholar]
  • 37. Cantineau JP, Lambert Y, Merckx P, Reynaud P, Porte F, Bertrand C, Duvaldestin P. End‐tidal carbon dioxide during cardiopulmonary resuscitation in humans presenting mostly with asystole: a predictor of outcome. Crit Care Med. 1996;24:791–796. [DOI] [PubMed] [Google Scholar]
  • 38. Axelsson C, Karlsson T, Axelsson AB, Herlitz J. Mechanical active compression‐decompression cardiopulmonary resuscitation (ACD‐CPR) versus manual CPR according to pressure of end tidal carbon dioxide (P(ET)CO2) during CPR in out‐of‐hospital cardiac arrest (OHCA). Resuscitation. 2009;80:1099–1103. [DOI] [PubMed] [Google Scholar]
  • 39. Hamrick JL, Hamrick JT, Lee JK, Lee BH, Koehler RC, Shaffner DH. Efficacy of chest compressions directed by end‐tidal CO2 feedback in a pediatric resuscitation model of basic life support. J Am Heart Assoc. 2014;3:e000450 DOI: 10.1161/JAHA.113.000450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Pytte M, Kramer‐Johansen J, Eilevstjønn J, Eriksen M, Strømme TA, Godang K, Wik L, Steen PA, Sunde K. Haemodynamic effects of adrenaline (epinephrine) depend on chest compression quality during cardiopulmonary resuscitation in pigs. Resuscitation. 2006;71:369–378. [DOI] [PubMed] [Google Scholar]
  • 41. Chung TN, Kim SW, Cho YS, Chung SP, Park I, Kim SH. Effect of vehicle speed on the quality of closed‐chest compression during ambulance transport. Resuscitation. 2010;81:841–847. [DOI] [PubMed] [Google Scholar]
  • 42. Roosa JR, Vadeboncoeur TF, Dommer PB, Panchal AR, Venuti M, Smith G, Silver A, Mullins M, Spaite D, Bobrow BJ. CPR variability during ground ambulance transport of patients in cardiac arrest. Resuscitation. 2013;84:592–595. [DOI] [PubMed] [Google Scholar]
  • 43. Hung SC, Mou CY, Hung HC, Lin IH, Lai SW, Huang JY. Chest compression fraction in ambulance while transporting patients with out‐of‐hospital cardiac arrest to the hospital in rural Taiwan. Emerg Med J. 2017;34:398–401. [DOI] [PubMed] [Google Scholar]
  • 44. Stapleton ER. Comparing CPR during ambulance transport: manual vs. mechanical methods. JEMS. 1991;16:63–64. [PubMed] [Google Scholar]
  • 45. Klouche K, Weil MH, Sun S, Tang W, Povoas HP, Kamohara T, Bisera J. Evolution of the stone heart after prolonged cardiac arrest. Chest. 2002;122:1006–1011. [DOI] [PubMed] [Google Scholar]
  • 46. Babini G, Grassi L, Russo I, Novelli D, Boccardo A, Luciani A, Fumagalli F, Staszewsky L, Fiordaliso F, De Maglie M, Salio M, Zani DD, Letizia T, Masson S, Luini MV, Pravettoni D, Scanziani E, Latini R, Ristagno G. Duration of untreated cardiac arrest and clinical relevance of animal experiments: the relationship between the “No‐Flow” duration and the severity of post‐cardiac arrest syndrome in a porcine model. Shock. 2018;49:205–212. [DOI] [PubMed] [Google Scholar]
  • 47. Foo NP, Chang JH, Su SB, Lin HJ, Chen KT, Cheng CF, Lin TY, Chen PC, Guo HR. A stabilization device to improve the quality of cardiopulmonary resuscitation during ambulance transportation: a randomized crossover trial. Resuscitation. 2013;84:1579–1584. [DOI] [PubMed] [Google Scholar]

Articles from Journal of the American Heart Association: Cardiovascular and Cerebrovascular Disease are provided here courtesy of Wiley

RESOURCES