Table 3.

Laboratory Testing Instrument Malfunctions and Their Impact on Study Protocol Testing, KCMC-Duke Health Collaboration Sections of KCRI Biotechnology Laboratory, November 2012 to October 2014

Section No. of Validated Instruments Total No. of Malfunction Frequencies Malfunction-Related Test Suspension Frequency, No. (%) Time to Engineer On-Site, Median (Range), d Time to Resolution, Median (Range), d
Chemistry 2 10 2 (20.0) 6 (1-29) 6 (1-136)
Hematology 2 8 2 (25.0) 16 (5-55) 25 (11-158)
Immunology 2 2 0 (0.0) 9 (8-9) 9 (8-9)
Microbiology 1 0 NA NA NA
Molecular 1 3 3 (100.0) 16 (6-30) 16 (6-30)
Total 8 23 7 (30.4) 9 (1-55) 9 (1-158)

KCMC, Kilimanjaro Christian Medical Centre; KCRI, Kilimanjaro Clinical Research Institute; NA, not applicable.