Table 1. Spectral accuracy and precision of IMS system as measured against Ocean Optics USB2G38810. Ocean Optics λOO –central wavelength as measured by Ocean Optics USB2G38810; Ocean Optics FWHM – full width at half maximum of laser line as measured by the Ocean Optics spectrometer; IMS λIMS – central wavelength of laser line as measured by the IMS system; IMS ± σ – standard deviation of measured central wavelength of the laser line as measured by the IMS;IMS FWHM – full width at half maximum of the laser line as measured by the IMS system; Δλ – difference between IMS and Ocean Optics central wavelength.

Ocean Optics
λOO[nm]
Ocean Optics
FWHM [nm]
IMS
λIMS[nm]
IMS ± 1σ
[nm]
IMS
FWHM [nm]
Δλ
λOOIMS [nm]
542 8.9 541.7 0.8 15.2 0.3
573.2 4.9 571.7 0.4 13.5 1.5
598.8 5.3 597.6 0.6 16.4 1.2
622 5.4 622.7 0.8 19.1 −0.7
651.7 6.4 652.4 1.0 23.3 −0.7
680.9 6.4 681.4 1.3 25.9 −0.5
698.1 7.7 697.9 1.3 26.9 0.2
703.6 7.8 702.1 2.1 29.9 1.5
723.1 8.8 719.7 3.6 30.0 3.4