Table 1. Spectral accuracy and precision of IMS system as measured against Ocean Optics USB2G38810. Ocean Optics λOO –central wavelength as measured by Ocean Optics USB2G38810; Ocean Optics FWHM – full width at half maximum of laser line as measured by the Ocean Optics spectrometer; IMS λIMS – central wavelength of laser line as measured by the IMS system; IMS ± σ – standard deviation of measured central wavelength of the laser line as measured by the IMS;IMS FWHM – full width at half maximum of the laser line as measured by the IMS system; Δλ – difference between IMS and Ocean Optics central wavelength.
Ocean Optics λOO[nm] |
Ocean Optics FWHM [nm] |
IMS λIMS[nm] |
IMS ± 1σ [nm] |
IMS FWHM [nm] |
Δλ λOO-λIMS [nm] |
---|---|---|---|---|---|
542 | 8.9 | 541.7 | 0.8 | 15.2 | 0.3 |
573.2 | 4.9 | 571.7 | 0.4 | 13.5 | 1.5 |
598.8 | 5.3 | 597.6 | 0.6 | 16.4 | 1.2 |
622 | 5.4 | 622.7 | 0.8 | 19.1 | −0.7 |
651.7 | 6.4 | 652.4 | 1.0 | 23.3 | −0.7 |
680.9 | 6.4 | 681.4 | 1.3 | 25.9 | −0.5 |
698.1 | 7.7 | 697.9 | 1.3 | 26.9 | 0.2 |
703.6 | 7.8 | 702.1 | 2.1 | 29.9 | 1.5 |
723.1 | 8.8 | 719.7 | 3.6 | 30.0 | 3.4 |