Structure delineation and image quality comparison between CTD-diffusion-weighted imaging and multiple instantaneous switchable scan-diffusion-weighted imaging by reader 2

Content Sequences Score of 3 Score of 2 Score of 1 Ratio of clearly structure delineation (%) Z P
Delineation of structure
 Zonal anatomy CTD-DWI 49 4 0 92.5 −1.000 0.317
MISS-DWI 50 3 0 94.3
 Prostate capsule CTD-DWI 48 5 0 90.6 −1.000 0.317
MISS-DWI 50 3 0 94.3
 Nerve vascular bundle CTD-DWI 43 10 0 84.9 −1.000 0.317
MISS-DWI 43 9 1 81.1
 Seminal vesicles CTD-DWI 43 10 0 84.9 −1.000 0.317
MISS-DWI 43 9 1 81.1
Image quality
 Sharpness CTD-DWI 47 6 0 88.7 0 1.000
MISS-DWI 47 6 0 88.7
 Contrast CTD-DWI 44 9 0 83.0 0 1.000
MISS-DWI 44 9 0 83.0
 Deformation CTD-DWI 47 6 0 88.7 −0.333 0.739
MISS-DWI 45 7 1 84.9
 Artifact CTD-DWI 41 12 0 77.4 −1.000 0.317
MISS-DWI 40 13 0 75.5