Table 1.

Parameters of MRI sequences for prostate RT Planning

T2W acquisition (2D T2W FSE) T2*W acquisition (2D “medic”)
FOV readout (phase) 240 mm (100%) 240 mm (100%)
PE oversampling 60% 60%
Number of Slices 28 28
Slice thickness/gap 2.5 mm/0 2.5 mm/0
Acquisition matrix (phase) 320 (75%) 256 (75%)
TE/TR 110 ms/7210 ms 24 ms/550 ms
Averages 3 2
Orientation Transaxial Transaxial
PE direction Left/right Left/right
Reconstruction matrix 320 × 320 512 × 512
Receiver bandwidth 200 Hz/pixel
Fat‐water shift = 0.84 mm
230 Hz/pixel
Fat‐water shift = 0.92 mm
Pixel size 0.75 mm × 0.75 mm 0.46875 mm × 0.46875 mm
Other Echo‐train length 25, echo spacing 9.98 ms, echo‐trains per slice 16 Combined echoes 5, flip Angle 28 degrees
Filters PrescanNormalize/DistCorrection 2D PrescanNormalize/DistCorrection 2D
Coil arrangement Spine coil & body array Spine coil & body array
Total acquisition time 2 min 46 s
Parallel imaging = 2 (GRAPPA)
6 min 4 s
Parallel imaging = 2 (GRAPPA)

FSE: fast spin echo; FOV: field of view; TE: echo time; TR: relaxation time; GRAPPA: GeneRalized Autocalibrating Partial Parallel Acquisition.