Skip to main content
. 2019 Feb 20;20(3):71–80. doi: 10.1002/acm2.12547

Table 1.

Measurement devices with assessed beam dimensions, their manufacturers, purpose, and nominal tolerances used during this study

Device (X × Y cm2) Manufacturer Purpose Nominal Tolerances
Daily QA 3 (20 × 20 cm2) Sun Nuclear Corporation, Melbourne, USA Output, symmetry ±3%
30013 Farmer ionization chamber (IC) in Solid Water (10 × 10 cm2) PTW, Freiburg, Germany Output ±2%
SNC Machine FS‐QA (15 × 15 cm2) Sun Nuclear Corporation, Melbourne, USA Symmetry ±2%
MPC (13.3 × 13.3 cm2) Varian Medical Systems, Palo Alto, CA, USA Output, uniformity ±2%