Skip to main content
. Author manuscript; available in PMC: 2020 May 1.
Published in final edited form as: Magn Reson Med. 2019 Jan 23;81(5):3108–3123. doi: 10.1002/mrm.27638

Figure 4.

Figure 4.

Results of undersampled (R=48) verification experiments, plotting parameter map error (σRi) per ROI as a function of the calculated aliasing noise quality factor (1/qi) for the three tested sequences (θbest, θaverage, and θworst). In (a), all data points are included achieving goodness of fit R2=0.89 and R2=0.20 in σR1 and σR2, respectively. In (b), obvious outliers from the σR2 are excluded, resulting in goodness of fit R2=0.87.